BACKGROUND

Remote cardiac monitoring and screening have already become an integral telemedicine component. The wide usage of several different wireless electrocardiography (ECG) devices warrants a validation study on their accuracy and reliability.

METHODS

Totally, 300 inpatients with the Nabz Hooshmand-1 handheld ECG device and the GE MAC 1200 ECG system (as the reference) were studied to check the accuracy of the devices in 1 and 6-limb lead performance. Simultaneous 10-second resting ECGs were assessed for the most common ECG parameters in lead I. Afterward, 6-lead ECGs (limb leads), were performed immediately and studied for their morphologies.

RESULTS

Of the 300 patients, 297 had acceptable ECG quality in both devices for simultaneous lead I ECGs. The ECGs were inspected on-screen by a cardiologist for their rhythms, rates, axes, numbers, morphologies of premature atrial and ventricular beats, morphologies and amplitudes of PQRST waves, P-wave durations, QRS-wave durations, P-R intervals, and QT intervals. No significant differences were detected between the devices, and no major abnormalities were missed. Six-limb lead ECGs were obtained in 284 patients, of whom 281 had acceptable quality in ECGs by both devices. The morphology matching evaluation of the ECGs demonstrated an overall 98% compatibility rate, with the highest compatibility in lead I and the lowest in lead augmented vector foot.

CONCLUSIONS

The diagnosis of critical pathological rhythms, including atrial fibrillation and high-grade atrioventricular node block, was not missed by the Nabz Hooshmand-1 and GE MAC 1200 ECG devices. Accordingly, rhythm detection as the primary purpose of handheld ECG devices was highly accurate. Both devices had acceptable sensitivity to diagnose long P-R and long and short QT intervals. Although the modern technology of smartphones and the physical inability for the 6-limb mode might cause old patients difficulty in utilizing such devices, their use for screening and follow-up is safe.

Для цитирования: Ahmadi-Renani Sajjad, Gharebaghi Milad, Kamalian Erfan, Hajghassem Hassan, Ghanbari Abolfazl, Karimi Alireza, Mansoury Bahman, Dayari Mohammad Saeed, Khatmi Nemati Mahdi, Karimi Armin, Zarghami Mohammad Hosein, Vasheghani-Farahani Ali. Clinical Validation of a Smartphone-based Handheld ECG Device: A Validation Study.. Critical pathways in cardiology. 2022. DOI: 10.1097/hpc.0000000000000303.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.