A Change of Pace: Modifying Pacemaker Settings in the Emergency Department as Emergency Physicians, A Case Report.
Аннотация
BACKGROUND
Pacemakers are implantable or external devices that send electrical impulses, allowing the heart to contract. Some common indications for permanent pacemaker placement include symptomatic bradycardia due to high-grade atrioventricular (AV) block, sick sinus syndrome, chronic bifascicular block, and postcardiac transplantation. Like all devices, pacemakers can malfunction or stop working entirely. We present a patient with an implantable pacemaker/cardioverter and defibrillator (ICD) displaying symptomatic failure to capture, and settings were adjusted in the emergency department (ED). Adjustment in the ED resolved the patient's symptoms and required no further interventions.
CASE REPORT: An 82-year-old woman with atrial fibrillation, heart failure, and a Medtronic dual chamber pacer/ICD (Medtronic, Minneapolis, Minnesota) presented to the ED for complaints of weakness and lightheadedness. Emergency medical services noted intermittent bradycardia, hypotension, and decreased responsiveness. Electrocardiogram (ECG) revealed failure to capture. The Medtronic application was used to interrogate and reprogram the device. The settings were DDD at a rate of 60 beats/min with ventricular output of 4.5 millivolts (mV) and atrial output of 4.5 mV, which were changed to ventricular lead outputs from 4.5 mV to 8.0 mV and rate from 60 to 80 for additional cardiac output. After this adjustment, the bradycardia resolved, and ECG showed a successful AV-paced rhythm. Transcutaneous pacing was never required. WHY SHOULD AN EMERGENCY PHYSICIAN BE AWARE OF THIS?: Adjustment of pacer settings in the ED to stabilize the critically ill patient is within the scope of the emergency physician. We can reduce stress to the patient and prevent hospitalizations. Pacemakers can become a tool to optimize treatment plans for an aging population in whom pacemakers are becoming more common.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.