MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.23486

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2013

Objective quantification of aortic valvular structures by cardiac computed tomography angiography in patients considered for transcatheter aortic valve implantation.

Korosoglou Grigorios, Gitsioudis Gitsios, Waechter-Stehle Irina, Weese Juergen, Krumsdorf Ulrike, Chorianopoulos Emmanuel, Hosch Waldemar, Kauczor Hans-Ulrich, Katus Hugo A, Bekeredjian Raffi

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To test the ability of a model-based segmentation of the aortic root for consistent assessment of aortic valve structures in patients considered for transcatheter aortic valve implantation (TAVI) who underwent 256-slice cardiac computed tomography (CT).

METHODS

Consecutive patients (n = 49) with symptomatic severe aortic stenosis considered for TAVI and patients without aortic stenosis (n = 17) underwent cardiac CT. Images were evaluated by two independent observers who measured the diameter of the aortic annulus and its distance to both coronary ostia (1) manually and (2) software-assisted. All acquired measures were compared with each other and to (3) fully automatic quantification.

RESULTS

High correlations were observed for 3D measures of the aortic annulus conducted on multiple oblique planes (r = 0.87 and 0.84 between observers and model-based measures, and r = 0.81 between observers). Reproducibility was further improved by software-assisted versus manual assessment for all the acquired variables (r = 0.98 versus 0.81 for annulus diameter, r = 0.94 versus 0.85 for distance to the left coronary ostium, P < 0.01 for both). Thus, using software-assisted measurements very low limits of agreement were observed for the annulus diameter (95%CI of -1.2 to 0.6 mm) and within very low time-spent (0.6 ± 0.1 min for software-assisted versus 1.6 ± 0.3 min per patient for manual assessment, P < 0.001). Assessment of the aortic annulus using the 3D model-based instead of manual 2D-coronal measurements would have modified the implantation strategy in 12 of 49 patients (25%) with aortic stenosis. Four of 12 patients with potentially modified implantation strategy yielded postprocedural moderate paravalvular regurgitation, which may have been avoided by implantation of a larger prosthesis, as suggested by automatic 3D measures.

CONCLUSION

Our study highlights the usefulness of software-assisted preprocedural assessment of the aortic annulus in patients considered for TAVI.

Для цитирования: Korosoglou Grigorios, Gitsioudis Gitsios, Waechter-Stehle Irina, Weese Juergen, Krumsdorf Ulrike, Chorianopoulos Emmanuel, Hosch Waldemar, Kauczor Hans-Ulrich, Katus Hugo A, Bekeredjian Raffi. Objective quantification of aortic valvular structures by cardiac computed tomography angiography in patients considered for transcatheter aortic valve implantation.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2013. DOI: 10.1002/ccd.23486.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.