DOI 10.1002/ccd.23370

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2012

Usefulness of transesophageal echocardiography during implantation of the CoreValve percutaneous aortic prosthesis: influence of the learning curve.

Cejudo Díaz del Campo Laura, Rubio Dolores Mesa, Ortiz Martin Ruiz, Ortega Mónica Delgado, Pan Manuel, Pineda Soledad Ojeda, Fernandez Elena Villanueva, Aguilera Jose Lopez, Suarez de Lezo Jose

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND

OBJECTIVES

The purpose of this study was to evaluate the usefulness of transesophageal echocardiography during the implantation of the CoreValve percutaneous aortic prosthesis and the influence of the learning curve in its indications.

METHODS

From April 2008 to January 2010, 53 patients were treated for symptomatic severe aortic stenosis with a CoreValve percutaneous aortic prosthesis. Transesophageal echocardiography was performed during the procedure for two reasons: if a discrepancy between the measures of the aortic annulus arose before the implantation and to provide visualization if a complication occurred during the procedure. Patients were divided in to two groups based on the learning curve (group I:the first 25 patients, group II: all of the subsequent patients).

RESULTS

Thirty-six transesophageal echocardiographies were performed on 28 patients (53%).In 15 cases the indication was to assist us in determining the proper sized prosthesis, and in the other 21 cases it was performed due to a complication during the procedure. The numbers of intraprocedure transesophageal echocardiographies were similar in both learning-curve groups, (64% in group I vs. 71% in group II, P = 0.56) but the proportion of the studies indicated by serious complications was significantly lower in the second period of the learning curve (67% in group I vs. 17% in group II, P < 0.05).

CONCLUSIONS

In our experience, transesophageal echocardiography was very useful in more than one-half of the patients during the implantation of the CoreValve percutaneous aortic prosthesis. After overcoming the learning curve, its indication due to serious procedure-related complications decreased significantly.

Для цитирования: Cejudo Díaz del Campo Laura, Rubio Dolores Mesa, Ortiz Martin Ruiz, Ortega Mónica Delgado, Pan Manuel, Pineda Soledad Ojeda, Fernandez Elena Villanueva, Aguilera Jose Lopez, Suarez de Lezo Jose. Usefulness of transesophageal echocardiography during implantation of the CoreValve percutaneous aortic prosthesis: influence of the learning curve.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2012. DOI: 10.1002/ccd.23370.