MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.athoracsur.2009.02.054

Медицинская библиотека · The Annals of thoracic surgery · 2009

Failure of percutaneous closure of prosthetic, aortic paravalvular leak.

Castedo Evaristo, Serrano-Fiz Santiago, Oteo Juan F, Ramis Sebastian, Martinez Paloma, Ugarte Juan

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

We report the progression of aortic insufficiency after percutaneous closure of an aortic prosthesis paravalvular leak with the Amplatzer vascular plug (AGA Inc, Golden Valley, MN). Removal of the device and replacement of the aortic prosthesis was successfully performed. Based on operative findings, we hypothesize that shape mismatch between the occluder system and the leak might promote tearing at the end of slanted defects further enhancing the regurgitant area.

Для цитирования: Castedo Evaristo, Serrano-Fiz Santiago, Oteo Juan F, Ramis Sebastian, Martinez Paloma, Ugarte Juan. Failure of percutaneous closure of prosthetic, aortic paravalvular leak.. The Annals of thoracic surgery. 2009. DOI: 10.1016/j.athoracsur.2009.02.054.