Mosnino E, Michel A-S, Messedi E, Kayem G, Pinton A
Аннотация
OBJECTIVE: To compare two portable lactate devices, Lactate Scout and StatStrip Xpress, to a laboratory reference device by assessing the concordance of cord blood lactate values at birth.
METHODS: We conducted a single-center prospective validation study in a level 3 maternity hospital. For all cord samples taken at birth, we analyzed lactates on the reference device (GEM4000®) and on two portable devices Lactate Scout and StatStrip Xpress. We compared the lactate values from each of the two handheld devices to the reference device.
RESULTS: A total of 101 samples were collected. Each was analyzed by all three machines. The average lactate value obtained by the different machines was 3.7mmol/L. The mean difference between the Lactate Scout and GEM4000® apparatus was 0.0mmol/L±0.8mmol/L. Between these two devices at lactate threshold of 4.8mmol/L, the negative predictive value was 96,2% (76/79) and the area under the curve was 0.98 (95% CI 0.96-1]. The mean difference between StatStrip Xpress and GEM4000® was 0.1mmol/L±0.7mmol/L. Between these two devices at the lactate threshold of 4.8mmol/L, the negative predictive value was 97,4% (76/78) and the area under the curve was 0.95 (95% CI 0.86-1).
CONCLUSION: There is a good correlation between the lactates obtained by the reference device and the two portable devices Lactate Scout and StatStrip Xpress.
METHODS: We conducted a single-center prospective validation study in a level 3 maternity hospital. For all cord samples taken at birth, we analyzed lactates on the reference device (GEM4000®) and on two portable devices Lactate Scout and StatStrip Xpress. We compared the lactate values from each of the two handheld devices to the reference device.
RESULTS: A total of 101 samples were collected. Each was analyzed by all three machines. The average lactate value obtained by the different machines was 3.7mmol/L. The mean difference between the Lactate Scout and GEM4000® apparatus was 0.0mmol/L±0.8mmol/L. Between these two devices at lactate threshold of 4.8mmol/L, the negative predictive value was 96,2% (76/79) and the area under the curve was 0.98 (95% CI 0.96-1]. The mean difference between StatStrip Xpress and GEM4000® was 0.1mmol/L±0.7mmol/L. Between these two devices at the lactate threshold of 4.8mmol/L, the negative predictive value was 97,4% (76/78) and the area under the curve was 0.95 (95% CI 0.86-1).
CONCLUSION: There is a good correlation between the lactates obtained by the reference device and the two portable devices Lactate Scout and StatStrip Xpress.