Use of infrared thermography to delineate temperature gradients and critical isotherms during catheter ablation with normal and half normal saline: Implications for safety and efficacy.
Аннотация
BACKGROUND
Radiofrequency (RF) ablation with half-normal saline (HNS) has shown promise as a bail-out strategy following failed ventricular tachycardia ablation using standard approaches.
OBJECTIVE
To use a novel infrared thermal imaging (ITI) model to evaluate biophysical and lesion characteristics during RF ablation using normal saline (NS) and HNS irrigation.
METHODS
Left ventricular strips of myocardium were excised from fresh porcine hearts. RF ablation was performed using an open-irrigated ablation catheter (Thermocool ST/SF) with NS (n = 75) and HNS (n = 75) irrigation using different power settings (40/50 W), RF durations (30/60 s), contact force of 10-15 g, and flow rate of 15 ml/min. RF lesions were recorded using an infrared thermal camera and border zone, lethal, 100° isotherms were matched with necrotic borders after 2% triphenyltetrazolium chloride staining. Lesion dimensions and isotherms (mm2 ) were measured.
RESULTS
In total, 150 lesions were delivered. HNS lesions were deeper (6.4 ± 1.1 vs. 5.7 ±0.8 mm; p = .03), and larger in volume (633 ± 153 vs. 468 ± 107 mm3 ; p = .007) than NS lesions. Steam pops (SPs) occurred during 19/75 lesions (25%) in the NS group and 32/75 lesions (43%) in the HNS group (p = .34). Lethal (57.8 ± 6.5 vs. 36.0 ± 3.9 mm2 ; p = .001) and 100°C isotherm areas (16.9 ± 6.9 vs. 3.8 ± 4.2 mm2 ; p = .003) areas were larger and were reached earlier in the HNS group.
CONCLUSIONS
RFA using HNS created larger lesions than NS irrigation but led to more frequent SPs. The presence of earlier lethal isotherms and temperature rises above 100°C on ITI suggest a potentially narrower therapeutic-safety window with HNS.