Experimental Validation and Calibration of Cherenkov Imaging for Real-Time Surface Dosimetry in Photon Breast Therapy.

Lu Baozhu, Zhu Yifeng, Chen Lisha, Teng Karen, Xiong Emily, Li Taoran, Freedman Gary, Taunk Neil, Pogue Brian W, Zhu Timothy C

СтатьяПеревод пока не подготовлен

Аннотация

Cherenkov imaging holds promise as a non-invasive, real-time method for surface dosimetry during photon-based external beam radiation therapy (EBRT). In this study, we evaluated the feasibility of using Cherenkov emission as a quantitative surrogate for surface dose in breast cancer treatment through Monte Carlo simulation and experimental validation. A Monte Carlo model was developed to simulate both dose deposition and corresponding Cherenkov photon emission for breast patients undergoing photon EBRT. Validation was performed using phantoms with homogeneous optical properties and varying geometries, enabling direct comparison between simulated and measured Cherenkov emission. Additionally, the simulated surface dose was validated against in-vivo measurements acquired using in-vivo dosimeters. Results show strong correlation between simulated and measured Cherenkov emission and surface dose, except at the field edges, where additional optical and geometric considerations are required. Based on these comparisons, a correction factor was derived to calibrate Cherenkov intensity to absolute surface dose. Finally, Cherenkov images acquired from breast cancer patients were compared with simulation results to assess the clinical utility of this approach. These findings support the use of Cherenkov imaging, when properly corrected, as a viable tool for real-time, patient-specific surface dosimetry in photon EBRT.

Для цитирования: Lu Baozhu, Zhu Yifeng, Chen Lisha, Teng Karen, Xiong Emily, Li Taoran, Freedman Gary, Taunk Neil, Pogue Brian W, Zhu Timothy C. Experimental Validation and Calibration of Cherenkov Imaging for Real-Time Surface Dosimetry in Photon Breast Therapy.. Proceedings of SPIE--the International Society for Optical Engineering. 2026. DOI: 10.1117/12.3081551.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.