Modeling Robotic-Assisted Mechanical Thrombectomy Procedures with the CorPath GRX Robot: The Core-Flow Study.

Tomasello Alejandro, Hernández David, Li Jiahui, Tiberi Riccardo, Rivera Eila, Vargas Joan Daniel, Losada Cristina, Jablonska Magda, Esteves Marielle, Diaz Maria Lourdes, Cendrero Judith, Requena Manuel, Diana Francesco, De Dios Marta, Singh Trisha, Gramegna Laura Ludovica, Ribo Marc

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND PURPOSE: Endovascular robotic devices may enable experienced neurointerventionalists to remotely perform endovascular thrombectomy. This study aimed to assess the feasibility, safety, and efficacy of robot-assisted endovascular thrombectomy compared with manual procedures by operators with varying levels of experience, using a 3D printed neurovascular model.

MATERIALS AND

METHODS

M1 MCA occlusions were simulated in a 3D printed neurovascular model, linked to a CorPath GRX robot in a biplane angiography suite. Four interventionalists performed manual endovascular thrombectomy (n = 45) and robot-assisted endovascular thrombectomy (n = 37) procedures. The outcomes included first-pass recanalization (TICI 2c-3), the number and size of generated distal emboli, and procedural length.

RESULTS

A total of 82 experimental endovascular thrombectomies were conducted. A nonsignificant trend favoring the robot-assisted endovascular thrombectomy was observed in terms of final recanalization (89.2% versus manual endovascular thrombectomy, 71.1%; P = .083). There were no differences in total mean emboli count (16.54 [SD, 15.15] versus 15.16 [SD, 16.43]; P = .303). However, a higher mean count of emboli of > 1 mm was observed in the robot-assisted endovascular thrombectomy group (1.08 [SD, 1.00] versus 0.49 [SD, 0.84]; P = .001) compared with manual endovascular thrombectomy. The mean procedural length was longer in robot-assisted endovascular thrombectomy (6.43 [SD, 1.71] minutes versus 3.98 [SD, 1.84] minutes; P < .001). Among established neurointerventionalists, previous experience with robotic procedures did not influence recanalization (95.8% were considered experienced; 76.9% were considered novices; P = .225).

CONCLUSIONS

In a 3D printed neurovascular model, robot-assisted endovascular thrombectomy has the potential to achieve recanalization rates comparable with those of manual endovascular thrombectomy within competitive procedural times. Optimization of the procedural setup is still required before implementation in clinical practice.

Для цитирования: Tomasello Alejandro, Hernández David, Li Jiahui, Tiberi Riccardo, Rivera Eila, Vargas Joan Daniel, Losada Cristina, Jablonska Magda, Esteves Marielle, Diaz Maria Lourdes, Cendrero Judith, Requena Manuel, Diana Francesco, De Dios Marta, Singh Trisha, Gramegna Laura Ludovica, Ribo Marc. Modeling Robotic-Assisted Mechanical Thrombectomy Procedures with the CorPath GRX Robot: The Core-Flow Study.. AJNR. American journal of neuroradiology. 2024. DOI: 10.3174/ajnr.a8205.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.