DOI 10.1007/s00464-022-09525-y
Feasibility of pressurized intra peritoneal aerosol chemotherapy using an ultrasound aerosol generator (usPIPAC).
Аннотация
BACKGROUND
We tested the feasibility of ultrasound technology for generating pressurized intraperitoneal aerosol chemotherapy (usPIPAC) and compared its performance vs. comparator (PIPAC).
MATERIAL AND
METHODS
A piezoelectric ultrasound aerosolizer (NextGen, Sinaptec) was compared with the available technology (Capnopen, Capnomed). Granulometry was measured for water, Glc 5%, and silicone oil using laser diffraction spectrometry. Two- and three-dimensional (2D and 3D) spraying patterns were determined with methylene blue. Tissue penetration of doxorubicin (DOX) was measured by fluorescence microscopy in the enhanced inverted Bovine Urinary Bladder model (eIBUB). Tissue DOX concentration was measured by high-performance liquid chromatography (HPLC).
RESULTS
The droplets median aerodynamic diameter was (usPIPAC vs. PIPAC): H20: 40.4 (CI 10-90%: 19.0-102.3) vs. 34.8 (22.8-52.7) µm; Glc 5%: 52.8 (22.2-132.1) vs. 39.0 (23.7-65.2) µm; Silicone oil: 178.7 (55.7-501.8) vs. 43.0 (20.2-78.5) µm. 2D and 3D blue ink distribution pattern of usPIPAC was largely equivalent with PIPAC, as was DOX tissue concentration (usPIPAC: 0.65 (CI 5-95%: 0.44-0.86) vs. PIPAC: 0.88 (0.59-1.17) ng/ml, p = 0.29). DOX tissue penetration with usPIPAC was inferior to PIPAC: usPIPAC: 60.1 (CI 5.95%: 58.8-61.5) µm vs. PIPAC: 1172 (1157-1198) µm, p < 0.001). The homogeneity of spatial distribution (top, middle and bottom of the eIBUB) was comparable between modalities.
DISCUSSION: usPIPAC is feasible, but its performance as a drug delivery system remains currently inferior to PIPAC, in particular for lipophilic solutions.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.