AI-Assisted 3D Intracardiac Echocardiography for Pulsed Field Ablation of Atrial Fibrillation Using a Novel Variable Loop Circular Catheter: A Multicenter Evaluation.

Dello Russo Antonio, Valeri Yari, Ciconte Giuseppe, Schiavone Marco, Compagnucci Paolo, Di Monaco Antonio, Riva Stefania, Salerno Raffaele, Volpato Giovanni, Cipolletta Laura, Parisi Quintino, Casella Michela, Grimaldi Massimo, Tondo Claudio, Pappone Carlo

СтатьяПеревод пока не подготовлен

Аннотация

Background

The VARIPULSE platform is an advanced Pulsed Field Ablation (PFA) system fully integrated with electro-anatomical mapping system, employing a variable loop circular catheter (VLCC) for atrial fibrillation (AF) ablation. The objective of the study is to assess for the first time the feasibility, safety, and procedural impact of AI (artificial intelligence)-assisted ICE (intracardiac echocardiography) mapping with the CARTOSOUND FAM Module compared with conventional electroanatomical mapping during PFA.

Methods

In this retrospective, multicenter study, 157 consecutive patients undergoing PFA for paroxysmal or persistent AF were included. Patients were divided into two groups: ICE-guided cohort (n = 64) and non-ICE-guided cohort (n = 93). Propensity score matching (PSM) was used to adjust for baseline differences.

Results

AI-assisted ICE mapping was feasible in all cases. Compared with conventional mapping, it significantly reduced LA (left atrium) mapping time (median 5 vs. 8 min; p < 0.001), LA dwell time (33.5 vs. 38.5 min; p = 0.001), and fluoroscopy time (7.5 vs. 14 min; p < 0.001). The total number of PFA applications was similar across groups (p = 0.136). No major adverse events occurred in either cohort during the procedure or within the first month of follow-up.

Conclusions

AI-assisted ICE mapping using the CARTOSOUND FAM Module enables accurate anatomical reconstruction and significantly optimizes procedural efficiency in PFA. This approach supports further development toward radiation-sparing and potentially fluoroscopy-free PFA workflows. For the first time, it addresses a gap in the current evidence regarding the use of ICE in PFA, building on evidence already established for radiofrequency ablation procedures.

Для цитирования: Dello Russo Antonio, Valeri Yari, Ciconte Giuseppe, Schiavone Marco, Compagnucci Paolo, Di Monaco Antonio, Riva Stefania, Salerno Raffaele, Volpato Giovanni, Cipolletta Laura, Parisi Quintino, Casella Michela, Grimaldi Massimo, Tondo Claudio, Pappone Carlo. AI-Assisted 3D Intracardiac Echocardiography for Pulsed Field Ablation of Atrial Fibrillation Using a Novel Variable Loop Circular Catheter: A Multicenter Evaluation.. Journal of clinical medicine. 2025. DOI: 10.3390/jcm14207249.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.