DOI 10.1016/j.vgie.2025.07.003
Metal biliary stent transection with over-the-scope clip removal system.
Аннотация
BACKGROUND AND AIMS: Uncovered self-expandable metal stents (UC-SEMSs) are used for unresectable pancreaticobiliary malignancies. These stents can fray, causing malfunction, ulceration, and perforation. Trimming frayed UC-SEMSs is challenging and risks injury to adjacent structures. Argon plasma coagulation (APC) has shown success transecting and shortening stent wires, but the technique can be cumbersome and time-consuming and risk tissue injury. No standardized power settings exist (65-85 W; 0.5-2 L/min). We present 2 cases of distorted UC-SEMSs requiring transection using a bipolar endoscopic system designed for over-the-scope clip (OTSC) removal. Electrosurgical generator costs are comparable (U.S. $3000-$7000), and single-use catheters cost U.S. $250-$500, making relative costs similar.
METHODS
Two distorted indwelling UC-SEMSs were transected with a bipolar OTSC removal device to accommodate further interventions. Electrical pulses fired only with adequate metal contact. The paired generator required several minutes of charging; alternative generators were incompatible.
RESULTS
In both cases, aberrant UC-SEMSs were successfully restructured using an OTSC removal system without adverse events or tissue injury.
CONCLUSIONS
APC can transect UC-SEMS wires, but these cases show a safe, effective, off-label method using an OTSC removal system.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.