DOI 10.1111/echo.15721

Медицинская библиотека · Echocardiography (Mount Kisco, N.Y.) · 2023

Performance of handheld ultrasound devices in diagnosis of pulmonary hypertension and right heart dysfunction in ICU patients.

Dayıoğlu Mürüvvet, Gürsel Gül, Özercan Seçil, Aydın Eda Macit, Nadastepe Özge

СтатьяПеревод пока не подготовлен

Аннотация

AIM: The use of handheld ultrasonography devices (HHUD) has increased recently but there are limited data about their performance in the evaluation of right heart dysfunction (RHD) and pulmonary hypertension (PHT). The aim of the study is to compare the performance of a HHUD with a conventional ultrasound device (CUD) in assessing PHT and RHD.

METHODS

This single-center prospective study was performed in a seven-bed teaching hospital intensive care unit (ICU). PHT and RHD criteria were compared by HHUD and CUD. Additional PHT criteria and right ventricle (RV) systolic and diastolic dysfunction criteria were also measured.

RESULTS

Forty-six patients were included in the study. There was no significant difference between the imaging rates and mean values of the parameters measured by both devices. When the positivity rates for additional PHT parameters and RHD criteria were compared, there were no significant differences between the devices. In Bland-Altman's analysis, there was good agreement and there was no bias between the measurements of the two devices but left ventricular end-systolic eccentricity index (LVSEI), right atrium area (RAA), and pulmonary artery diameter (PAD). Ninety percent of the patients had PHT probability, of whom 43% had a low, 37% had intermediate and 10% had a high probability of PHT. Ninety-two percent of the patients had RHD and there was no significant difference between the devices in the diagnosis of RHD (p = .212). When RV systolic and diastolic dysfunction evaluations of the devices were compared according to the British Society of Echocardiography (BSE) criteria there was no significant difference between the devices' measurements in the evaluation of systolic and diastolic function.

CONCLUSION

The imaging and measurement capabilities of the HHUDs for PHT and RHD parameters were similar to CUDs, and considering the inconsistent parameters, the HHUD can be useful in diagnosing these problems.

Для цитирования: Dayıoğlu Mürüvvet, Gürsel Gül, Özercan Seçil, Aydın Eda Macit, Nadastepe Özge. Performance of handheld ultrasound devices in diagnosis of pulmonary hypertension and right heart dysfunction in ICU patients.. Echocardiography (Mount Kisco, N.Y.). 2023. DOI: 10.1111/echo.15721.