Clinica chimica acta; international journal of clinical chemistry · 2021
Consideration triggered by the choice of heterophilic antibody interference detection tests in measuring ACTH for a teenager boy with a rare adrenal disease.
Клинический случайСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Yin Yicong, Zhao Fang, Hu Yingying, Lu Lin, Qiu Ling
Аннотация
BACKGROUND: Plasma ACTH is an important marker for Cushing syndrome; inaccurate measurement of ACTH concentrations due to the serum heterophile antibody may mislead diagnosis by physicians. We describe the case of a teenage boy with a rare adrenal disease who presented falsely increased ACTH concentrations (20.3 pg/ml; Siemens Immulite).
METHODS: Low dose dexamethasone inhibition test showed no inhibition of serum cortisol. While biochemical evidence suggested ACTH-dependent Cushing, the imaging tests suggested otherwise. Hence, further tests for interference identification, including test platform replacement, sample dilution, PEG precipitation, and use of blocking antibodies, were conducted.
RESULTS: The measurements of the same patient sample on three different analytical platforms suggested ACTH concentrations below 5 pg/ml. Serial dilutions of serum samples revealed nonlinearity, PEG recovery resulted in ACTH concentrations dropping to < 5 pg/ml, and interferences in plasma were blocked by a heterophilic blocking reagent (21.2 vs < 5 pg/ml), suggesting that assay interference performed on Siemens Immulite led to false-positive ACTH. Finally, the boy was diagnosed with a rare type of ACTH-independent Cushing syndrome.
CONCLUSIONS: The inaccurate evaluation of ACTH misled the physicians, increasing treatment time and diagnostic testing. Thus, an anti-interference workflow should be established between physicians and laboratories to avoid such cases in the future.
METHODS: Low dose dexamethasone inhibition test showed no inhibition of serum cortisol. While biochemical evidence suggested ACTH-dependent Cushing, the imaging tests suggested otherwise. Hence, further tests for interference identification, including test platform replacement, sample dilution, PEG precipitation, and use of blocking antibodies, were conducted.
RESULTS: The measurements of the same patient sample on three different analytical platforms suggested ACTH concentrations below 5 pg/ml. Serial dilutions of serum samples revealed nonlinearity, PEG recovery resulted in ACTH concentrations dropping to < 5 pg/ml, and interferences in plasma were blocked by a heterophilic blocking reagent (21.2 vs < 5 pg/ml), suggesting that assay interference performed on Siemens Immulite led to false-positive ACTH. Finally, the boy was diagnosed with a rare type of ACTH-independent Cushing syndrome.
CONCLUSIONS: The inaccurate evaluation of ACTH misled the physicians, increasing treatment time and diagnostic testing. Thus, an anti-interference workflow should be established between physicians and laboratories to avoid such cases in the future.