Left atrial appendage closure: Initial experience with the ultraseal device.
Аннотация
OBJECTIVES
We report the initial experience of percutaneous left atrial appendage (LAA) closure with the Ultraseal device.
BACKGROUND
LAA closure is an alternative to oral anticoagulation for stroke prevention in patients with nonvalvular atrial fibrillation (NVAF) and high risk of bleeding. The Ultraseal device is a new LAA closure prosthesis that consists of a distal soft bulb and a proximal sail attached by an articulating joint that allows a high degree of device conformability to the different variations of the LAA anatomy.
METHODS
We included 12 consecutive patients with NVAF who underwent LAA closure with the Ultraseal device between January and December 2015 in our center. Patients had clinical and transesophageal echocardiography (TEE) examinations at baseline and at 45 days following LAA closure.
RESULTS
The device was successfully implanted in all patients, with no periprocedural complications. There were no episodes of bleeding, stroke, pericardial effusion, or device embolization at 45-day follow-up. No cases with residual leaks >5 mm were observed at TEE. One patient presented a device related thrombus without clinical consequences.
CONCLUSIONS
This initial experience with the Ultraseal LAA closure device demonstrates preliminary safety and feasibility. Further larger studies with longer follow-up are warranted. © 2016 Wiley Periodicals, Inc.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.