Quantitative biomechanical optimization of neochordal implantation location on mitral leaflets during valve repair.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Pandya Pearly K, Wilkerson Robert J, Imbrie-Moore Annabel M, Zhu Yuanjia, Marin-Cuartas Mateo, Park Matthew H, Woo Y Joseph
Аннотация
OBJECTIVE: Suture pull-out remains a significant mechanism of long-term neochordal repair failure, as demonstrated by clinical reports on recurrent mitral valve regurgitation and need for reoperation. The objective of this study was to provide a quantitative comparison of suture pull-out forces for various neochordal implantation locations.
METHODS: Posterior leaflets were excised from fresh porcine mitral valves (n = 54) and fixed between two 3-dimensional-printed plates. Gore-Tex CV-5 sutures (WL Gore & Associates Inc) were placed with distances from the leading edge and widths between anchoring sutures with values of 2 mm, 6 mm, and 10 mm for a total of 9 groups (n = 6 per group). Mechanical testing was performed using a tensile testing machine to evaluate pull-out force of the suture through the mitral valve leaflet.
RESULTS: Increasing the suture anchoring width improved failure strength significantly across all leading-edge distances (P < .001). Additionally, increasing the leading-edge distance from 2 mm to 6 mm increased suture pull-out forces significantly across all suture widths (P < .001). For 6-mm and 10-mm widths, increasing the leading-edge distance from 6 mm to 10 mm increased suture pull-out forces by an average of 3.58 ± 0.15 N; in comparison, for leading-edge distances of 6 mm and 10 mm, increasing the suture anchoring width from 6 mm to 10 mm improves the force by an average of 7.09 ± 0.44 N.
CONCLUSIONS: Increasing suture anchoring width and leading-edge distance improves the suture pull-out force through the mitral leaflet, which may optimize postrepair durability. The results suggest a comparative advantage to increasing suture anchoring width compared with leading-edge distance.
METHODS: Posterior leaflets were excised from fresh porcine mitral valves (n = 54) and fixed between two 3-dimensional-printed plates. Gore-Tex CV-5 sutures (WL Gore & Associates Inc) were placed with distances from the leading edge and widths between anchoring sutures with values of 2 mm, 6 mm, and 10 mm for a total of 9 groups (n = 6 per group). Mechanical testing was performed using a tensile testing machine to evaluate pull-out force of the suture through the mitral valve leaflet.
RESULTS: Increasing the suture anchoring width improved failure strength significantly across all leading-edge distances (P < .001). Additionally, increasing the leading-edge distance from 2 mm to 6 mm increased suture pull-out forces significantly across all suture widths (P < .001). For 6-mm and 10-mm widths, increasing the leading-edge distance from 6 mm to 10 mm increased suture pull-out forces by an average of 3.58 ± 0.15 N; in comparison, for leading-edge distances of 6 mm and 10 mm, increasing the suture anchoring width from 6 mm to 10 mm improves the force by an average of 7.09 ± 0.44 N.
CONCLUSIONS: Increasing suture anchoring width and leading-edge distance improves the suture pull-out force through the mitral leaflet, which may optimize postrepair durability. The results suggest a comparative advantage to increasing suture anchoring width compared with leading-edge distance.