Mechanical Cardiac Support With an Implantable Direct Cardiac Compression Device: Proof of Concept.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Letsou George V, Bolch Christina M, Hord Erica C, Altman William C, Youngblood Brad L, Leschinsky Boris, Criscione John C
Аннотация
PURPOSE: We examined the hemodynamic effects of a new, implantable, direct cardiac assist device in an ovine heart failure model.
DESCRIPTION: The device, which encompasses both left and right ventricles, is inserted through the pericardial apex and self-expands to encompass the heart without suturing. The intact pericardium anchors the device in place. The device has 2 concentric chamber layers: an internal chamber layer filled with fluid to conform to the heart and an external chamber layer filled with air that provides external compression and negative pressure to aid relaxation.
EVALUATION: The device was implanted in 7 sheep with heart failure induced by microsphere embolization. Cardiac performance was assessed for 6 to 8 hours. The cardiac assist device provided cardiac systolic and diastolic assistance, as shown by pressure tracings of the left ventricle and aorta, pulmonary artery flow, and +dP/dt. Central venous pressure decreased during cardiac assistance. No anatomic damage was noted postmortem.
CONCLUSIONS: Systolic and diastolic cardiac assistance can be achieved with this device that compresses and relaxes in synchrony with the native cardiac cycle.
DESCRIPTION: The device, which encompasses both left and right ventricles, is inserted through the pericardial apex and self-expands to encompass the heart without suturing. The intact pericardium anchors the device in place. The device has 2 concentric chamber layers: an internal chamber layer filled with fluid to conform to the heart and an external chamber layer filled with air that provides external compression and negative pressure to aid relaxation.
EVALUATION: The device was implanted in 7 sheep with heart failure induced by microsphere embolization. Cardiac performance was assessed for 6 to 8 hours. The cardiac assist device provided cardiac systolic and diastolic assistance, as shown by pressure tracings of the left ventricle and aorta, pulmonary artery flow, and +dP/dt. Central venous pressure decreased during cardiac assistance. No anatomic damage was noted postmortem.
CONCLUSIONS: Systolic and diastolic cardiac assistance can be achieved with this device that compresses and relaxes in synchrony with the native cardiac cycle.