Novel stent design for transcatheter mitral valve implantation.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Loger Klaas, Pokorny Saskia, Schaller Tim, Haben Irma, Frank Derk, Lutter Georg
Аннотация
OBJECTIVES: In this study, results of a functional in vitro study of 2 newly developed valved stents for transcatheter mitral valve implantation are presented.
METHODS: Two novel stent designs, an oval-shaped and a D-shaped stent with a strut fixation system were developed. The fixation force of the novel stents were tested in vitro in porcine hearts with a tensile test set-up. In further experiments, the stents were equipped with a circular valved stent, and the valve performances were investigated in a pulsatile heart valve tester.
RESULTS: Sufficient mean stent fixation forces in the range of 24.2 ± 0.9 N to 28.6 ± 1.9 N were measured for the different stent models. The novel valved stents showed good performance in an in vitro pulsatile heart valve tester. A sufficient opening area and low opening pressures were measured for all tested mitral valved stents. Compared with an established reference valve, the D-shaped stent and the oval-shaped valved stent showed a lower systolic transvalvular pressure gradient, which indicates slightly greater extent of valvular leakage of the closed valved stents. However, the mitral nitinol valved stents demonstrated adequate durability.
CONCLUSIONS: This study indicates a sufficient annular fixation force of the tested transcatheter mitral valve implantation valved stent prototypes. Therefore, these mitral valved stents demonstrate a new type of mitral valved stent design.
METHODS: Two novel stent designs, an oval-shaped and a D-shaped stent with a strut fixation system were developed. The fixation force of the novel stents were tested in vitro in porcine hearts with a tensile test set-up. In further experiments, the stents were equipped with a circular valved stent, and the valve performances were investigated in a pulsatile heart valve tester.
RESULTS: Sufficient mean stent fixation forces in the range of 24.2 ± 0.9 N to 28.6 ± 1.9 N were measured for the different stent models. The novel valved stents showed good performance in an in vitro pulsatile heart valve tester. A sufficient opening area and low opening pressures were measured for all tested mitral valved stents. Compared with an established reference valve, the D-shaped stent and the oval-shaped valved stent showed a lower systolic transvalvular pressure gradient, which indicates slightly greater extent of valvular leakage of the closed valved stents. However, the mitral nitinol valved stents demonstrated adequate durability.
CONCLUSIONS: This study indicates a sufficient annular fixation force of the tested transcatheter mitral valve implantation valved stent prototypes. Therefore, these mitral valved stents demonstrate a new type of mitral valved stent design.