MedElevator Войти в исследования
← Поиск публикаций

DOI 10.23736/s0021-9509.17.10065-0

Медицинская библиотека · The Journal of cardiovascular surgery · 2017

Politetrafluorene suture used as artificial mitral chord: mechanical properties and surgical implications.

Caimmi Philippe P, Sabbatini Maurizio, Fusaro Luca, Cannas Mario

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Novel surgical approach to repair degenerative mitral regurgitation such as transapical chordae tendineae replacement and "loop in loop" in loop techniques, need of artificial chordae longer than that used in the older techniques of chordae tendineae replacement. This difference in length has been reported as potential critical point for durability of artificial chordae. In the present paper we have investigated the elastic behavior of different diameter and length politetrafluorene (PTFE) suture threads as substitute of native chordae, to identify their reliability to use as long artificial chordae.

METHODS

PTFE suture threads with different diameters were investigated in their mechanical properties at different length from 2 to 14 cm, by a servo hydraulic testing machine, to test the elastic properties of the sample in their use as mitral chordae substitutes.

RESULTS

Our study shows that the chordae length is an important parameter that can change the performance of chordae itself. The analysis of elastic/properties of suture threads specimen, reveals that long PTFE chords have an optimal mechanical behavior in which elongation is accompanied by a safe elastic properties that make them well resistance during multiple tractions.

CONCLUSIONS

In conclusion the use of PTFE as an artificial chordae may represent a valid choice in case of insertion of artificial chordae with extra anatomic length.

Для цитирования: Caimmi Philippe P, Sabbatini Maurizio, Fusaro Luca, Cannas Mario. Politetrafluorene suture used as artificial mitral chord: mechanical properties and surgical implications.. The Journal of cardiovascular surgery. 2017. DOI: 10.23736/s0021-9509.17.10065-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.