Argon plasma coagulation in the right and left colon: safety-risk profile of the 60W-1.2 l/min setting.

Panos Marios Z, Koumi Andriani

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND STUDY AIM: The 40W-0.8 l/min setting is widely recommended for argon plasma coagulation (APC) in the right colon. Until March 2012, we used the 60W-1.2 l/min setting for all sites of the colon. By auditing our experience, we assessed the safety-risk profile of the 60W-1.2 l/min setting in the right and left colon.

PATIENTS AND

METHODS

All cases treated with APC by a single endoscopist, using the 60W-1.2 l/min setting for all sites of the colon between October 2001 and December 2007 were identified retrospectively and site, type, number of lesions, and complications were recorded. Between January 2008 and March 2012, information was recorded prospectively.

RESULTS

In the retrospective audit, 290 lesions (101 cecum/ascending, 120 sigmoid/descending, 69 transverse) were treated in 241 patient endoscopies. There were no perforations. In the prospective audit, 156 lesions (83 cecum/ascending, 47 sigmoid/descending, 26 transverse) were treated in 132 patient endoscopies. There was 1/83 (1.2%) perforation in the cecum/ascending colon and none in the transverse or sigmoid/descending (n.s.). Combined, the results yield a cecal/ascending perforation rate of 1/153 (0.6%) patient endoscopies, 1/184 (0.5%) lesions treated and overall perforation rate for all sites of the colon of 1/373 (0.3%) patient endoscopies and 1/446 (0.2%) lesions. Post-polypectomy syndrome and delayed bleeding each occurred in 3/373 (0.8%) patient endoscopies and 3/446 (0.7%) lesions. There were no deaths.

CONCLUSION

In the cecum and ascending colon, the APC perforation rate at the 60W-1.2 l/min setting was no higher than in the left colon and is similar to that reported in previously published series. Therefore, it appears safe, provided the precautions we describe are strictly followed.

Для цитирования: Panos Marios Z, Koumi Andriani. Argon plasma coagulation in the right and left colon: safety-risk profile of the 60W-1.2 l/min setting.. Scandinavian journal of gastroenterology. 2014. DOI: 10.3109/00365521.2014.903510.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.