DOI 10.1016/j.hrthm.2017.04.024

Медицинская библиотека · Heart rhythm · 2017

Risk of atrioesophageal fistula formation with contact force-sensing catheters.

Black-Maier Eric, Pokorney Sean D, Barnett Adam S, Zeitler Emily P, Sun Albert Y, Jackson Kevin P, Bahnson Tristram D, Daubert James P, Piccini Jonathan P

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Atrioesophageal fistula formation is a rare but life-threatening complication of atrial fibrillation ablation. Contact force (CF)-sensing catheters improve procedural effectiveness. However, the impact of the implementation of CF-sensing technology on the risk of atrioesophageal fistula formation has not been explored.

OBJECTIVE

The purpose of this study was to determine the association between the use of CF-sensing catheters and atrioesophageal fistula development.

METHODS

We searched the Manufacturer and User Facility Device Experience database for adverse event reports involving Food and Drug Administration-approved ablation catheters.

RESULTS

Among 2689 device reports, we identified 78 atrioesophageal fistula cases, 65 of which involved CF-sensing catheters and 13 non-CF-sensing catheters. The percentage of total reports involving atrioeosphageal fistula was 5.4% for CF-sensing catheters (65 of 1202) and 0.9% for non-CF-sensing catheters (13 of 1487) (P < .0001). Procedural details (CF and power settings) were not consistently reported. Esophageal temperature increases were detected in only 2.5% of cases (2 of 78). The mean time to presentation was 16 ± 9 days. Overall mortality was at least 56%, with patients who underwent surgical repair more likely to survive than those treated with stenting or no intervention.

CONCLUSION

Atrioesophageal fistula formation accounted for a much higher proportion of reported adverse events with CF-sensing catheters compared with non-CF-sensing catheters. Improved understanding of the relationship between power/force delivery and esophageal damage is needed to minimize the risk of atrioesophageal fistula formation.

Для цитирования: Black-Maier Eric, Pokorney Sean D, Barnett Adam S, Zeitler Emily P, Sun Albert Y, Jackson Kevin P, Bahnson Tristram D, Daubert James P, Piccini Jonathan P. Risk of atrioesophageal fistula formation with contact force-sensing catheters.. Heart rhythm. 2017. DOI: 10.1016/j.hrthm.2017.04.024.

Предметные рубрики MeSH