DOI 10.4244/eijv11i5a109

Медицинская библиотека · EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology · 2015

Comparison of neointimal coverage between everolimus-eluting stents and sirolimus-eluting stents: an optical coherence tomography substudy of the RESET (Randomized Evaluation of Sirolimus-eluting versus Everolimus-eluting stent Trial).

Kubo Takashi, Akasaka Takashi, Kozuma Ken, Kimura Kazuo, Kawamura Masaki, Sumiyoshi Tetsuya, Ino Yasushi, Morino Yoshihiro, Tanabe Kengo, Kadota Kazushige, Kimura Takeshi, RESET Investigators

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: The aim of the present study was to compare vascular healing response between the first-generation sirolimus-eluting stent (SES) and the second-generation everolimus-eluting stent (EES) by using optical coherence tomography (OCT).

METHODS AND

RESULTS

The RESET was a prospective, multicentre, randomised, open-label trial comparing EES and SES. Of the 3,197 patients enrolled in the RESET, nine-month follow-up OCT after stent implantation was performed in 100 patients (48 EES-treated lesions in 44 patients and 62 SES-treated lesions in 56 patients), thus constituting the OCT substudy population. The percentage of uncovered struts per lesion (8±15% vs. 14±19%, p=0.031) and cross-sections with >30% uncovered struts per lesion (6±14% vs. 18±29%, p=0.009) was significantly lower in EES compared with SES. The frequency of DES-treated lesions with incomplete stent apposition (8 [17%] vs. 29 [47%], p<0.001) was significantly lower in EES compared with SES. Intra-stent thrombus was comparably observed between EES and SES (1 [2%] vs. 5 [8%], p=0.229).

CONCLUSIONS

Compared with SES, EES was associated with a favourable vascular response at nine months after stent implantation.

Для цитирования: Kubo Takashi, Akasaka Takashi, Kozuma Ken, Kimura Kazuo, Kawamura Masaki, Sumiyoshi Tetsuya, Ino Yasushi, Morino Yoshihiro, Tanabe Kengo, Kadota Kazushige, Kimura Takeshi, RESET Investigators. Comparison of neointimal coverage between everolimus-eluting stents and sirolimus-eluting stents: an optical coherence tomography substudy of the RESET (Randomized Evaluation of Sirolimus-eluting versus Everolimus-eluting stent Trial).. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2015. DOI: 10.4244/eijv11i5a109.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.