MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.rec.2022.03.003

Медицинская библиотека · Revista espanola de cardiologia (English ed.) · 2022

Feasibility of precise commissural and coronary alignment with balloon-expandable TAVI.

Santos-Martínez Sandra, Redondo Alfredo, González-Bartol Esther, Barrero Alejandro, Sánchez-Luna Juan Pablo, Revilla-Orodea Ana, Baladrón Carlos, Serrador Ana, San Román J Alberto, Amat-Santos Ignacio J

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION AND

OBJECTIVES

We aimed to describe the feasibility and preliminary outcomes of commissural alignment (CA) for the balloon-expandable transcatheter heart valve.

METHODS

The relationship among native commissures and transcatheter aortic valve implantation neocommissures was analyzed in 10 consecutive patients with tricuspid severe aortic stenosis undergoing transcatheter aortic valve implantation after guided implantation based on computed tomography analysis with a self-developed software. CA was predicted by in silico bio-modelling in the 10 patients and the calculated rotation was applied during crimping. Degrees of CA and coronary overlap (CO) were measured through 1-month follow up computed tomography. Transvalvular residual gradients and the rate of paravalvular leak were also analyzed.

RESULTS

Mean commissural misalignment was 16.7±8°. Four patients showed mild misalignment but none of them showed a moderate or severe degree of misalignment. The in silico model accurately predicted the final in vivo position with a correlation coefficient of 0.983 (95%CI, 0.966-0.992), P <.001. Severe CO with right coronary ostium occurred in 3 patients likely due to ostial eccentricity, and CO was not present with the left coronary artery in any of the patients. Mean transaortic gradient was 6.1±3.3mmHg and there were no moderate-severe paravalvular leaks.

CONCLUSIONS

Patient-specific rotation during valve crimping based on preprocedural computed tomography is feasible with balloon-expandable devices and is associated with the absence of moderate or severe commissural misalignment and left main CO.

Для цитирования: Santos-Martínez Sandra, Redondo Alfredo, González-Bartol Esther, Barrero Alejandro, Sánchez-Luna Juan Pablo, Revilla-Orodea Ana, Baladrón Carlos, Serrador Ana, San Román J Alberto, Amat-Santos Ignacio J. Feasibility of precise commissural and coronary alignment with balloon-expandable TAVI.. Revista espanola de cardiologia (English ed.). 2022. DOI: 10.1016/j.rec.2022.03.003.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.