Efficacy of larger valve sizing with underfilling in balloon-expandable transcatheter aortic valve replacement.

Kawaguchi Tomohiro, Miyoshi Toru, Hayashi Masaomi, Ishizu Kenichi, Ugawa Satoko, Nosaka Kazumasa, Isotani Akihiro, Doi Masayuki, Shirai Shinichi, Ando Kenji

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To investigate the efficacy and safety of larger valve sizing beyond the commercially recommended annular range in transcatheter aortic valve replacement (TAVR) with balloon-expandable transcatheter heart valve (THVs).

BACKGROUND

The clinical implications of larger balloon-expandable THV implantation with underfilling are poorly evaluated.

METHODS

This retrospective study included 692 consecutive patients who underwent TAVR with SAPIEN3. A total of 271 patients who underwent SAPIEN 3 implantation were analyzed based on three border zones (Zone 1: 300-345 mm2 , 23 vs. 20 mm; Zone 2: 400-430 mm2 , 26 vs. 23 mm; Zone 3: 500-546 mm2 , 29 vs. 26 mm). The primary endpoint was the effective orifice area (EOA) assessed by echocardiography at 1 year, and secondary endpoints were a 30-day mortality rate, procedural complications during TAVR, and a composite of death from any cause and heart failure requiring rehospitalization at 1 year.

RESULTS

At 1-year follow-up, the EOA in the larger valve groups was greater than that in the recommended valve group in each zone (Zone 1: 1.45 ± 0.03 vs. 1.06 ± 0.06 cm2 , p < 0.001; Zone 2: 1.83 ± 0.05 vs. 1.41 ± 0.05 cm2 , p < 0.001; Zone 3: 1.93 ± 0.07 vs. 1.69 ± 0.07 cm2 , p = 0.02). No significant difference in the secondary endpoint was observed in any of the zones.

CONCLUSIONS

Implantation of the out-of-range larger SAPIEN 3 THVs with underfilling was associated with greater EOA at the 1-year follow-up and feasible in the selected patients.

Для цитирования: Kawaguchi Tomohiro, Miyoshi Toru, Hayashi Masaomi, Ishizu Kenichi, Ugawa Satoko, Nosaka Kazumasa, Isotani Akihiro, Doi Masayuki, Shirai Shinichi, Ando Kenji. Efficacy of larger valve sizing with underfilling in balloon-expandable transcatheter aortic valve replacement.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2022. DOI: 10.1002/ccd.30188.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.