DOI 10.1080/02656736.2024.2405105

Медицинская библиотека · International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group · 2024

Evaluation of magnetic resonance thermometry performance during MR-guided hyperthermia treatment of soft-tissue sarcomas in the lower extremities and pelvis.

Karkavitsas Spyridon N, Göger-Neff Marianne, Kawula Maria, Sumser Kemal, Zilles Benjamin, Wadepohl Martin, Landry Guillaume, Kurz Christopher, Kunz Wolfgang G, Dietrich Olaf, Lindner Lars H, Paulides Margarethus M

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

This study evaluated the performance of magnetic resonance thermometry (MRT) during deep-regional hyperthermia (HT) in pelvic and lower-extremity soft-tissue sarcomas.

MATERIALS AND

METHODS

17 pelvic (45 treatments) and 16 lower-extremity (42 treatments) patients underwent standard regional HT and chemotherapy. Pairs of double-echo gradient-echo scans were acquired during the MR protocol 1.4 s apart. For each pair, precision was quantified using phase data from both echoes ('dual-echo') or only one ('single-echo') in- or excluding body fat pixels in the field drift correction region of interest. The precision of each method was compared to that of the MRT approach using a built-in clinical software tool (SigmaVision). Accuracy was assessed in three lower-extremity patients (six treatments) using interstitial temperature probes. The Jaccard coefficient quantified pretreatment motion; receiver operating characteristic analysis assessed its predictability for acceptable precision (<1 °C) during HT.

RESULTS

Compared to the built-in dual-echo approach, single-echo thermometry improved the mean temporal precision from 1.32 ± 0.40 °C to 1.07 ± 0.34 °C (pelvis) and from 0.99 ± 0.28 °C to 0.76 ± 0.23 °C (lower extremities). With body fat-based field drift correction, single-echo mean accuracy improved from 1.4 °C to 1.0 °C. Pretreatment bulk motion provided excellent precision prediction with an area under the curve of 0.80-0.86 (pelvis) and 0.81-0.83 (lower extremities), compared to gastrointestinal air motion (0.52-0.58).

CONCLUSION

Single-echo MRT exhibited better precision than dual-echo MRT. Body fat-based field-drift correction significantly improved MRT accuracy. Pretreatment bulk motion showed improved prediction of acceptable MRT temporal precision over gastrointestinal air motion.

Для цитирования: Karkavitsas Spyridon N, Göger-Neff Marianne, Kawula Maria, Sumser Kemal, Zilles Benjamin, Wadepohl Martin, Landry Guillaume, Kurz Christopher, Kunz Wolfgang G, Dietrich Olaf, Lindner Lars H, Paulides Margarethus M. Evaluation of magnetic resonance thermometry performance during MR-guided hyperthermia treatment of soft-tissue sarcomas in the lower extremities and pelvis.. International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group. 2024. DOI: 10.1080/02656736.2024.2405105.