Development of an external system for monitoring the couch speed in radiotherapy using continuous bed movement.

Shimizu Hidetoshi, Nakamura Osamu, Sasaki Koji, Aoyama Takahiro, Kitagawa Tomoki, Kodaira Takeshi

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Total body irradiation before bone marrow transplantation for hematological malignancies using Radixact, a high-precision radiotherapy machine, can potentially reduce side effects and the risk of secondary malignancies. However, stable control of couch speed is critical, and direct assessment methods outlined in quality assurance guidelines are lacking. This study aims to develop a real-time couch speed verification system for the Radixact.

METHODS

The developed system used a linear encoder to measure couch speed directly. Accuracy was verified via a linear stage, comparing measurements with a laser distance sensor. After placing a phantom simulating the human body on the Radixact couch, the couch speed was verified using predefined speed plans.

RESULTS

Operating the linear stage at 0.1, 0.5, and 1.0 mm/s revealed that the maximum position error of the developed verification system compared to the laser distance sensor was nearly equivalent to the distance resolution of the system (0.05 mm/pulse), with negligible average speed error. When the Radixact couch operated at 0.1, 0.5, and 1.0 mm/s, the values obtained by the verification system agreed with the theoretical values within the sampling period (0.01 s) and distance resolution (0.05 mm).

CONCLUSION

The verification system developed provides real-time monitoring of the speed of the Radixact table, ensuring treatment effectiveness and patient safety. It would guarantee the couch speed's soundness and contribute to the "visualization" of safety.

Для цитирования: Shimizu Hidetoshi, Nakamura Osamu, Sasaki Koji, Aoyama Takahiro, Kitagawa Tomoki, Kodaira Takeshi. Development of an external system for monitoring the couch speed in radiotherapy using continuous bed movement.. Journal of applied clinical medical physics. 2024. DOI: 10.1002/acm2.14497.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.