BACKGROUND

Ablation of atrial tachyarrhythmia in adults with congenital heart disease (ACHD) is challenging because of complex anatomy and high scar burden. We proposed that the addition of high-density mapping with the PentaRay® (Biosense Webster, Inc) mapping catheter (EAM+P) to 3-dimensional electroanatomic mapping (EAM) allows for rapid acquisition of high-resolution maps and shorter procedure times.

METHODS

In this single-center, retrospective cohort study of patients with ACHD who underwent atrial arrhythmia ablation, patients were divided those who underwent ablation with EAM and those who underwent ablation with EAM+P.

RESULTS

Fifteen ablations were performed in 13 patients using standard EAM, and 11 ablations were performed in 10 patients using EAM+P. There was no difference in mean age or complexity of congenital heart disease. The procedure duration was 1.5 times longer in the EAM than in the EAM+P group (P = .015). The dose area product was 12 times higher in the EAM than in the EAM+P group (P = .001). A higher number of venous access sites were used for EAM cases than for EAM+P cases (P = .008). Acute success rates of ablation and recurrence rates at 1 year were similar in the 2 groups. There were no procedure-related complications in either group.

CONCLUSION

This is the first study to evaluate the use of the PentaRay® high-density mapping catheter for ablation of atrial tachyarrhythmia in patients with ACHD. The use of the PentaRay® high-density mapping catheter results in shorter procedure time, decreased radiation exposure, and fewer venous access sites.

Для цитирования: Dodeja Anudeep K, Tan Yubo, Ackley Tamara, Russell Jennifer, Kertesz Naomi, Daniels Curt J, Kamp Anna. PentaRay® Multielectrode Mapping Catheter for Atrial Tachyarrhythmia in Adults With Congenital Heart Disease.. Texas Heart Institute journal. 2022. DOI: 10.14503/thij-20-7535.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.