MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/mrm.30256

Медицинская библиотека · Magnetic resonance in medicine · 2024

Abdominal MR Multitasking for radiotherapy treatment planning: A motion-resolved and multicontrast 3D imaging approach.

Chen Junzhou, Christodoulou Anthony G, Han Pei, Xiao Jiayu, Han Fei, Hu Zhehao, Wang Nan, Han Hui, Ling Diane C, Chang Eric L, Feng Mary, Scholey Jessica E, Cui Sophia, Li Debiao, Yang Wensha, Fan Zhaoyang

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: Radiotherapy treatment planning (RTP) using MR has been used increasingly for the abdominal site. Multiple contrast weightings and motion-resolved imaging are desired for accurate delineation of the target and various organs-at-risk and patient-tailored planning. Current MR protocols achieve these through multiple scans with distinct contrast and variable respiratory motion management strategies and acquisition parameters, leading to a complex and inaccurate planning process. This study presents a standalone MR Multitasking (MT)-based technique to produce volumetric, motion-resolved, multicontrast images for abdominal radiotherapy treatment planning.

METHODS

The MT technique resolves motion and provides a wide range of contrast weightings by repeating a magnetization-prepared (saturation recovery and T2 preparations) spoiled gradient-echo readout series and adopting the MT image reconstruction framework. The performance of the technique was assessed through digital phantom simulations and in vivo studies of both healthy volunteers and patients with liver tumors.

RESULTS

In the digital phantom study, the MT technique presented structural details and motion in excellent agreement with the digital ground truth. The in vivo studies showed that the motion range was highly correlated (R2 = 0.82) between MT and 2D cine imaging. MT allowed for a flexible contrast-weighting selection for better visualization. Initial clinical testing with interobserver analysis demonstrated acceptable target delineation quality (Dice coefficient = 0.85 ± 0.05, Hausdorff distance = 3.3 ± 0.72 mm).

CONCLUSION

The developed MT-based, abdomen-dedicated technique is capable of providing motion-resolved, multicontrast volumetric images in a single scan, which may facilitate abdominal radiotherapy treatment planning.

Для цитирования: Chen Junzhou, Christodoulou Anthony G, Han Pei, Xiao Jiayu, Han Fei, Hu Zhehao, Wang Nan, Han Hui, Ling Diane C, Chang Eric L, Feng Mary, Scholey Jessica E, Cui Sophia, Li Debiao, Yang Wensha, Fan Zhaoyang. Abdominal MR Multitasking for radiotherapy treatment planning: A motion-resolved and multicontrast 3D imaging approach.. Magnetic resonance in medicine. 2024. DOI: 10.1002/mrm.30256.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.