Cardiac responses of vagus nerve stimulation: intraoperative bradycardia and subsequent chronic stimulation.

Ardesch J J, Buschman H P J, van der Burgh P H, Wagener-Schimmel L J J C, van der Aa H E, Hageman G

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Few adverse events on heart rate have been reported with vagus nerve stimulation (VNS) for refractory epilepsy. We describe three cases with intraoperative bradycardia during device testing.

PATIENTS AND

METHODS

At our hospital 111 patients have received a VNS system. Intraoperative device testing is performed under ECG-monitoring. We reviewed the patients and their VNS-therapy follow-up outcome who experienced a change in heart rate, during device testing (Lead Test).

RESULTS

Three patients with medically refractory epilepsy showed a bradycardia during intraoperative Lead Test. Postoperative the VNS-therapy started under ECG-monitoring. No change in cardiac rhythm occurred. Subsequent chronic stimulation is uneventful. All three have reduced seizure frequency. Two already have had their second implant, without the occurrence of bradycardia.

CONCLUSION

In case of intraoperative bradycardia VNS-therapy onset should be done under ECG-monitoring. Subsequent chronic stimulation is safe in respect to heart rate. Bradycardia during intraoperative device testing is no reason to abort the operation.

Для цитирования: Ardesch J J, Buschman H P J, van der Burgh P H, Wagener-Schimmel L J J C, van der Aa H E, Hageman G. Cardiac responses of vagus nerve stimulation: intraoperative bradycardia and subsequent chronic stimulation.. Clinical neurology and neurosurgery. 2007. DOI: 10.1016/j.clineuro.2007.07.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.