A new technique of direct intra-abdominal pressure measurement: a preliminary study.

Risin Elena, Kessel Boris, Ashkenazi Itamar, Lieberman Noah, Alfici Ricardo

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

The purpose of this study was to determine if a 14-F polyvinyl chloride (PVC) round drain is a reliable tool for direct intra-abdominal pressure measurement.

DESIGN

A prospective interventional study.

SETTING

Department of Surgery B, intensive care unit, recovery room, Hillel-Yaffe level II trauma center.

METHODS

Forty patients undergoing abdominal surgery and treated postoperatively with intraperitoneal drains and intravesical catheters were included in this study. The indication for insertion of intraperitoneal drains and intravesical catheters was strictly medical. The decision of placing urinary bladder catheter and PVC round drain was done by a senior surgeon. Intra-abdominal pressures were measured simultaneously through the intraperitoneal drain and the urinary catheter. Using a sterile technique, the intraperitoneal drain was disconnected from the drainage bulb and connected to an invasive blood pressure monitoring system. Intravesical pressures were measured by inserting 50 mL into the bladder, and then the urinary catheter was connected to an invasive blood monitoring system. Measurements were done twice a day for 3 days or less if earlier removal of either the intraperitoneal drain or urinary catheters were medically indicated.

RESULTS

Two hundred twenty-even simultaneous measurements were performed. Pressures as measured through the intraperitoneal drain were found to be significantly correlated to pressures as measured intravesically (r = 0.962).

CONCLUSIONS

Direct measurement of the intra-abdominal pressure via a 14-F PVC round drain is a newly described technique. Our method is simple, safe, and credible. Future investigation is needed to confirm the reliability of this method for continuous postoperative measurement of the intra-abdominal pressure in selected patients.

Для цитирования: Risin Elena, Kessel Boris, Ashkenazi Itamar, Lieberman Noah, Alfici Ricardo. A new technique of direct intra-abdominal pressure measurement: a preliminary study.. American journal of surgery. 2006. DOI: 10.1016/j.amjsurg.2005.07.038.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.