MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.urology.2005.07.007

Медицинская библиотека · Urology · 2006

Irrigation and drainage properties of three-way urethral catheters.

Braasch Matthew, Antolak Christopher, Hendlin Kari, Botnaru Andrei, Herrera Shannon, Lavers Ann, Canales Benjy, Monga Manoj

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

To evaluate in an in vitro study the irrigation and drainage capabilities of a variety of available catheters. Three-way Foley catheters are used when irrigation of the bladder is anticipated to prevent or manage blood clots in the bladder.

METHODS

Six operators applied maximal one-handed pressure irrigation with a catheter-tip syringe through the irrigation and drainage ports of the Bardex and Dover three-way catheters (16F to 26F). Maximal manual irrigation flow was recorded using a flowmeter. Continuous gravity irrigation and drainage flows were measured by suspending a 2-L irrigation bag 80 cm above the catheters and fastening an artificial bladder to the drainage tip of each Foley catheter with the catheter balloon inflated to 30 mL.

RESULTS

The Bardex 22F and 24F catheters had superior manual irrigation using the irrigation port (maximal urinary flow rate 8.9 +/- 1.3 mL/s and 9.5 +/- 1.5 mL/s, respectively, P < 0.000001) compared with the Dover 22F and 24F catheters. The Bardex and Dover 22F catheters (maximal urinary flow rate 29.0 +/- 1.9 mL/s and 29.3 +/- 1.9 mL/s, respectively, P = 0.66) and 24F catheters (maximal urinary flow rate 30.5 +/- 2.0 mL/s and 30.6 +/- 1.2 mL/s, respectively, P = 0.83) performed equally in manual irrigation using the drainage port. The Bardex 22F (average flow rate 1.6 +/- 0.1 mL/s, P < 0.00001) and 24F (average flow rate 1.7 +/- 0.0 mL/s, P < 0.000001) catheters exhibited better continuous irrigation and flow than the Dover 22F and 24F catheters.

CONCLUSIONS

Three-way catheters are placed in situations in which efficient bladder irrigation is essential for patient safety. The Bardex 22F to 26F catheters optimized continuous bladder irrigation of the catheters tested.

Для цитирования: Braasch Matthew, Antolak Christopher, Hendlin Kari, Botnaru Andrei, Herrera Shannon, Lavers Ann, Canales Benjy, Monga Manoj. Irrigation and drainage properties of three-way urethral catheters.. Urology. 2006. DOI: 10.1016/j.urology.2005.07.007.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.