Use of the inverse solution guidance algorithm method for RF ablation catheter guidance.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Lv Wener, Barrett Conor D, Arai Tatsuya, Bapat Aneesh, Armoundas Antonis A, Cohen Richard J, Lee Kichang
Аннотация
INTRODUCTION: We previously introduced the inverse solution guidance algorithm (ISGA) methodology using a Single Equivalent Moving Dipole model of cardiac electrical activity to localize both the exit site of a re-entrant circuit and the tip of a radiofrequency (RF) ablation catheter. The purpose of this study was to investigate the use of ISGA for ablation catheter guidance in an animal model.
METHODS: Ventricular tachycardia (VT) was simulated by rapid ventricular pacing at a target site in eleven Yorkshire swine. The ablation target was established using three different techniques: a pacing lead placed into the ventricular wall at the mid-myocardial level (Type-1), an intracardiac mapping catheter (Type-2), and an RF ablation catheter placed at a random position on the endocardial surface (Type-3). In each experiment, one operator placed the catheter/pacing lead at the target location, while another used the ISGA system to manipulate the RF ablation catheter starting from a random ventricular location to locate the target.
RESULTS: The average localization error of the RF ablation catheter tip was 0.31 ± 0.08 cm. After analyzing approximately 35 cardiac cycles of simulated VT, the ISGA system's accuracy in locating the target was 0.4 cm after four catheter movements in the Type-1 experiment, 0.48 cm after six movements in the Type-2 experiment, and 0.67 cm after seven movements in the Type-3 experiment.
CONCLUSION: We demonstrated the feasibility of using the ISGA method to guide an ablation catheter to the origin of a VT focus by analyzing a few beats of body surface potentials without electro-anatomic mapping.
METHODS: Ventricular tachycardia (VT) was simulated by rapid ventricular pacing at a target site in eleven Yorkshire swine. The ablation target was established using three different techniques: a pacing lead placed into the ventricular wall at the mid-myocardial level (Type-1), an intracardiac mapping catheter (Type-2), and an RF ablation catheter placed at a random position on the endocardial surface (Type-3). In each experiment, one operator placed the catheter/pacing lead at the target location, while another used the ISGA system to manipulate the RF ablation catheter starting from a random ventricular location to locate the target.
RESULTS: The average localization error of the RF ablation catheter tip was 0.31 ± 0.08 cm. After analyzing approximately 35 cardiac cycles of simulated VT, the ISGA system's accuracy in locating the target was 0.4 cm after four catheter movements in the Type-1 experiment, 0.48 cm after six movements in the Type-2 experiment, and 0.67 cm after seven movements in the Type-3 experiment.
CONCLUSION: We demonstrated the feasibility of using the ISGA method to guide an ablation catheter to the origin of a VT focus by analyzing a few beats of body surface potentials without electro-anatomic mapping.