BACKGROUND

Implantable cardioverter defibrillator (ICD) therapies, even when appropriate, are associated with increased risk. Therapy-reducing strategies have been shown to reduce the mortality rate.

METHODS AND

RESULTS

In total, 895 patients with ICD and cardiac resynchronization therapy with defibrillation function (CRT-D) were included in the study; of these, 506 (57%) patients undergoing secondary prevention were included. Devices implanted before May 2014 were programmed according to conventional programming (CP), the others according to our novel programming (NP) with high rate cut-off, longer detection intervals and 4-6 anti-tachycardia pacing (ATP) trains in the ventricular tachycardia (VT) zone. Time-to-first-event for mortality, appropriate and inappropriate therapies were analyzed. Follow-up time was 24.0 months (IQR 13.0-24.0 months). There was a significant reduction in mortality rate (11.4% vs. 25.4%, P<0.001) and in the rate of appropriate (18.8% vs. 42.2%, P<0.001) and inappropriate therapies (5.2% vs. 18.0%, P<0.001) with NP according to Kaplan-Meier analyses. In multivariate analysis, NP (hazard ratio [HR]=0.35; P<0.001), chronic kidney disease (HR=1.55), reduced ejection fraction (EF) (HR=1.35), secondary ICD indication (HR=2.35) and age at implantation (HR=1.02) were associated with mortality reduction. NP was also associated with significant reduction in the rate of appropriate and inappropriate therapies. These results were consistent after stratification for primary and secondary prevention.

CONCLUSIONS

Novel ICD programming reduced mortality and morbidity due to appropriate or inappropriate ICD therapies in secondary as well as in primary ICD indication.

Для цитирования: Schober Alexander Daniel, Schober Andreas Leonhard, Hubauer Ute, Fredersdorf Sabine, Seegers Joachim, Keyser Andreas, Stadler Stefan, Birner Christoph, Maier Lars, Jungbauer Carsten, Ücer Ekrem. Novel Implantable Cardioverter Defibrillator Programming With High Rate Cut-Off, Long Detection Intervals and Multiple Anti-Tachycardia Pacing Reduces Mortality.. Circulation journal : official journal of the Japanese Circulation Society. 2021. DOI: 10.1253/circj.cj-20-0940.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.