Journal of vascular surgery · 2016

Experience with the GORE EXCLUDER iliac branch endoprosthesis for common iliac artery aneurysms.

СтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

van Sterkenburg Steven M M, Heyligers Jan M M, van Bladel Mathijs, Verhagen Hence J, Eefting Daniël, van Sambeek Marc R, Zeebregts Clark J, Reijnen Michel M P J, Dutch IBE Collaboration

Аннотация

OBJECTIVE: In this study, we analyzed the procedural success and early outcome of endovascular treatment of a multicenter cohort of patients with common iliac artery (CIA) aneurysms treated with the new GORE EXCLUDER (W. L. Gore & Associates, Flagstaff, Ariz) iliac branch endoprosthesis (IBE).

METHODS: A retrospective cohort analysis was performed in 13 sites in The Netherlands. Anatomic, demographic, procedural, and follow-up data were assessed from hospital records.

RESULTS: From November 2013 to December 2014, 51 CIA aneurysms were treated with an IBE in 46 patients. The median diameter of the treated aneurysm was 40.5 (range, 25.0-90.0) mm. The mean procedural time was 198 ± 56 minutes. All but one implantation were successful; two type Ib endoleaks were noticed, resulting in a procedural success rate of 93.5%. The two type Ib endoleaks spontaneously disappeared at 30 days. There was no 30-day mortality. Ipsilateral buttock claudication was present in only two cases at 30 days and disappeared during follow-up. The incidence of reported erectile dysfunction was low and severe ischemic complications were absent. After a mean follow-up of 6 months, data on 17 treated aneurysms were available. Two showed a stable diameter, whereas 15 showed a mean decrease of 3.9 ± 2.2 mm (P < .001). Reinterventions were performed in two patients (7.1%). The 6-month primary patency of the internal component of the IBE device was 94%.

CONCLUSIONS: The use of the GORE EXCLUDER IBE device for CIA aneurysms is related to high procedural success, high patency rates, and low reintervention rates at short-term follow-up. Prospective data with longer follow-up are awaited to establish the role of the device in the treatment algorithm of CIA aneurysms.