MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/17085381231162128

Медицинская библиотека · Vascular · 2023

Treating post-angioplasty dissection in the femoropopliteal arteries using the tack endovascular system: Tack optimized balloon angioplasty II 24-month results.

Brodmann Marianne, Werner Martin, Sood Ami, Gray William A

СтатьяМногоцентровое исследованиеКлиническое исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

The TOBA (Tack Optimized Balloon Angioplasty) II trial is a prospective, single-arm, multicenter study that investigated Tack treatment for patients with dissection after angioplasty in the superficial femoral artery and/or proximal popliteal artery. The Tack device is a nitinol-based, short (6 mm), stent-like implant with low outward force that can be deployed in a targeted fashion to treat vascular dissection. TOBA II primary results through 12 months have been published previously. This report provides follow-up safety and efficacy results through 24 months (RC).

METHODS

The TOBA II trial enrolled 213 patients with Rutherford classification 2 to 4 and a de novo or non-stented restenotic lesion in the superficial femoral artery and/or proximal popliteal artery who developed a dissection of any grade after treatment with plain balloon or drug-coated balloon (DCB) angioplasty. Participants were followed for 30 days, 6 months, 12 months, 24 months, and 36 months following the procedure. Evaluations included clinically driven target lesion revascularization (CD-TLR), ankle-brachial index, Rutherford classification, peripheral artery questionnaire, quality of life assessed by the EQ-5D-3L, and the Walking Impairment Questionnaire.

RESULTS

At enrollment, mean age was 68.2 ± 9.1 years, 70.9% were male, and 95.8% of patients were categorized as RC 2 or 3. The distribution of balloon types in the study were standard balloons: 42.3%; and drug-coated balloons: 57.7%. At 24-month follow-up, 167 patients (78.4%) had available data. The overall survival rate at 24 months was 95.4% and there were no major amputations during this time. After 24 months of follow-up, the Kaplan-Meier freedom from CD-TLR was 77.7%. Rutherford classification, ankle-brachial index, and quality of life were significantly improved compared with baseline through 24 months.

CONCLUSIONS

The TOBA II 24-month data demonstrate durable intermediate-term outcomes with the use of the Tack Endovascular System. Tack deployment was a safe and effective therapeutic option for dissection repair following angioplasty.

Для цитирования: Brodmann Marianne, Werner Martin, Sood Ami, Gray William A. Treating post-angioplasty dissection in the femoropopliteal arteries using the tack endovascular system: Tack optimized balloon angioplasty II 24-month results.. Vascular. 2023. DOI: 10.1177/17085381231162128.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.