Initial experience with directed laser atherectomy using the CLiRpath photoablation atherectomy system and bias sheath in superficial femoral artery lesions.

Rastan Aljoscha, Sixt Sebastian, Schwarzwälder Uwe, Kerker William, Bürgelin Karlheinz, Frank Ulrich, Noory Elias, Gremmelmeier Daniel, Branzan Daniela, Hauswald Kirsten, Brantner Regina, Schwarz Thomas, Zeller Thomas

Клиническое исследование I фазыСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To report a safety and efficacy study of directed excimer laser atherectomy with the Bias Sheath guiding catheter to create larger lumens in the superficial femoral artery (SFA).

METHODS

Fourteen patients (9 men; mean age 66+/-9 years, range 46-76) with chronic lower limb ischemia (Rutherford class 2 to 5) referable to SFA stenoses were treated with the 8-F-compatible Bias Sheath and a 1.7- or 2.0-mm laser catheter. Eight (57%) lesions were de novo and 6 (43%) were in-stent restenoses (ISR). Mean diameter stenosis was 96%+/-8% (range 80%-100%; 10 total occlusions), and mean lesion length was 126+/-78 mm (range 30-290). The primary study endpoint was laser success, defined as <50% angiographic residual stenosis post-laser, without major perforations.

RESULTS

The primary endpoint was achieved in 8 (57%) lesions; all patients underwent additional balloon angioplasty, which achieved an angiographic residual stenosis <30% in all. Three lesions could not be completely traversed with the study device. Two dissections were detected after Bias Sheath application; 1 required stenting. There were 2 embolic events (specific to the antegrade device design); both were treated with aspiration therapy. Intravascular ultrasound (IVUS)-based minimal lumen diameters were 0.23+/-0.2 mm (range 0-1.0) at baseline and 3.2+/-0.8 mm after Bias Sheath application. Rutherford category improved at 1 month in 10 (71%) patients. One patient remained unchanged, another worsened (category 2 to 3), 1 died, and the fourth was not examined. The ankle-brachial index improved from 0.41+/-0.18 at baseline to 0.79+/-0.19 at discharge and to 0.80+/-0.19 at 1 month. Primary 1-month patency (<50% restenosis by duplex) was 85% (11/13 lesions).

CONCLUSION

SFA lesions, including ISR, can be treated in the majority of cases with directed laser atherectomy, significantly reducing plaque burden as measured by IVUS. Embolization was attributed to device-specific features of the prototype antegrade sheath design, which was discontinued. Larger studies are mandatory to document the long-term technical and clinical impact of this new device.

Для цитирования: Rastan Aljoscha, Sixt Sebastian, Schwarzwälder Uwe, Kerker William, Bürgelin Karlheinz, Frank Ulrich, Noory Elias, Gremmelmeier Daniel, Branzan Daniela, Hauswald Kirsten, Brantner Regina, Schwarz Thomas, Zeller Thomas. Initial experience with directed laser atherectomy using the CLiRpath photoablation atherectomy system and bias sheath in superficial femoral artery lesions.. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2007. DOI: 10.1583/06-2046.1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.