Automated end-to-side anastomosis to the middle cerebral artery: a feasibility study.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Bregy Amadé, Alfieri Alex, Demertzis Stefanos, Mordasini Pasquale, Jetzer Anna Katharina, Kuhlen Dominique, Schaffner Thomas, Dacey Ralph, Steiger Hans-Jakob, Reinert Michael
Аннотация
OBJECT: The treatment of complex cerebrovascular or skull base pathological conditions necessitates a microsurgical blood flow preservation or augmentative revascularization procedure as either an adjunctive safety measure or a definitive treatment. The brain is susceptible to ischemia, and procedure-related risks can be minimized by the reduction of occlusion time or the use of a nonocclusive technique. The authors therefore analyzed the feasibility of an automatic device (C-Port xA, Cardica) designed for constructing an end-to-side anastomosis with or without flow interruption for a middle cerebral artery (MCA) bypass in a human cadaveric model and in an in vivo craniotomy simulation model.
METHODS: Four Thiel-fixated human head specimens were prepared using 8 standard pterional craniotomies. The sylvian fissure was opened to access the anterior circulation and in particular the MCA. The length of the individual vessel segments was measured. The C-Port xA was tested on each of the 8 exposures. In addition the C-Port xA was deployed in an in vivo craniotomy simulator model in 10 New Zealand rabbits (a total of 20 anastomoses) by using the abdominal aorta jump graft model.
RESULTS: Short-term patency was assessed by angiography and histological findings. In all 8 sylvian exposures, construction of an MCA anastomosis with the aid of the C-Port xA was feasible. All 20 jump graft anastomoses performed in the in vivo craniotomy simulator were found to be patent.
CONCLUSIONS: The anatomical studies as well as the in vivo craniotomy simulation studies demonstrated that the dimensions of the automated end-to-side anastomosis device are suitable for an extracranial-intracranial high-flow bypass on the MCA. Further miniaturization and special adaptation of this device would allow bypass procedures to more proximal intracranial vessels.
METHODS: Four Thiel-fixated human head specimens were prepared using 8 standard pterional craniotomies. The sylvian fissure was opened to access the anterior circulation and in particular the MCA. The length of the individual vessel segments was measured. The C-Port xA was tested on each of the 8 exposures. In addition the C-Port xA was deployed in an in vivo craniotomy simulator model in 10 New Zealand rabbits (a total of 20 anastomoses) by using the abdominal aorta jump graft model.
RESULTS: Short-term patency was assessed by angiography and histological findings. In all 8 sylvian exposures, construction of an MCA anastomosis with the aid of the C-Port xA was feasible. All 20 jump graft anastomoses performed in the in vivo craniotomy simulator were found to be patent.
CONCLUSIONS: The anatomical studies as well as the in vivo craniotomy simulation studies demonstrated that the dimensions of the automated end-to-side anastomosis device are suitable for an extracranial-intracranial high-flow bypass on the MCA. Further miniaturization and special adaptation of this device would allow bypass procedures to more proximal intracranial vessels.