A novel approach to calculate the required volume of air for bronchial blockers in young children.

Zhu Change, Zhang Saiji, Zhang Mazhong, Wei Rong

СтатьяПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

INTRODUCTION

Bronchial blocker balloons inflated with small volumes of air increase balloon pressure, involving a risk of airway injury especially in young children. However, there are no established guidelines regarding the appropriate volumes of air required to provide safe bronchial occlusion.

METHODS

This study aimed to introduce a novel method for calculating the amount of air required for safe bronchial blocker balloon occlusion for one lung anesthesia in young children. We included 79 pediatric patients who underwent video-assisted thoracoscopic surgery at our hospital. Preoperatively, the balloon pressure and corresponding diameter of 5F bronchial blockers inflated with different volumes of air were measured. Intraoperatively, bronchial diameters measured by computerized tomographic scans were matched to the ex vivo measured balloon diameters. The quality of lung isolation, incidence of balloon repositioning, and airway injury were documented. Postoperatively, airway injury was evaluated through fiberoptic bronchoscopy.

RESULTS

Balloon pressure and balloon diameter showed linear and nonlinear correlations with volume, respectively. The median lengths of the right and left mainstem bronchi were median (interquartile range) range: 5.3 mm (4.5-6.3) 2.7-8.15 and 21.8 (19.6-23.4) 14-29, respectively. Occluding the left mainstem bronchus required <1 mL of air, with a balloon pressure of 27 cm H2O. The isolation quality was high with no case of mucosal injury or displacement. Occluding the right mainstem bronchus required a median air volume of 1.3 mL, with a median balloon pressure of 44 cm H2O. One patient had poor lung isolation due to a tracheal bronchus and another developed mild and transient airway injury.

CONCLUSION

The bronchial blocker cuff should be regarded as a high-pressure balloon. We introduced a new concept for safe bronchial blocker balloon occlusion for one-lung ventilation in small children.

Для цитирования: Zhu Change, Zhang Saiji, Zhang Mazhong, Wei Rong. A novel approach to calculate the required volume of air for bronchial blockers in young children.. Paediatric anaesthesia. 2024. DOI: 10.1111/pan.14964.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.