Evaluation of a new free-thyroxin assay.

Wellby M L, Guthrie L, Reilly C P

СтатьяПеревод пока не подготовлен

Аннотация

The Amerlex Free Thyroxin (T4) Radioimmunoassay Kit (Amersham International Ltd.) is a new direct equilibrium radioimmunoassay for free T4 based on an antiserum with very high affinity for T4, and a unique 125I-labeled T4 analog as tracer. It is a very simple single-tube radioimmunoassay, making use of Amerlex particles to separate antibody-bound from free species. Interassay precision (CV) is 3.7% at 13 pmol/L and 2.3% at 30 pmol/L; within-assay precision is 4.2% at 21 pmol/L. The reference interval is 11-22 pmol/L. The assay did not misclassify any patients tested who had untreated myxedema or untreated thyrotoxicosis. The free T4 assay excelled both the free T4 index and the T4/T4-binding globulin ratio in correcting for increased thyroxin-binding globulin from pregnancy, and it was better than the index but not better than the ratio in correcting for increased thyroxin-binding globulin in users of oral contraceptives.