The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation · 2023

Ambulatory 7-day mechanical circulatory support in sheep model of pulmonary hypertension and right heart failure.

СтатьяПоддержка NIH: внешняя программаНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ukita Rei, Patel Yatrik J, Kelly Wu W, Francois Sean A, Cortelli Michael, Johnson Carl A, Cardwell Nancy, Talackine Jennifer R, Stokes John W, Grogan William, Mentz Meredith, Tracy Kaitlyn M, Harris Timothy R, Tucker William, Simonds Elizabeth, Demarest Caitlin T, Cook Keith E, Skoog David J, Rosenzweig Erika B, Bacchetta Matthew

Аннотация

BACKGROUND: Right heart failure is the major cause of death in pulmonary hypertension. Lung transplantation is the only long-term treatment option for patients who fail medical therapy. Due to the scarcity of donor lungs, there is a critical need to develop durable mechanical support for the failing right heart. A major design goal for durable support is to reduce the size and complexity of devices to facilitate ambulation. Toward this end, we sought to deploy wearable mechanical support technology in a sheep disease model of chronic right heart failure.

METHODS: In 6 sheep with chronic right heart failure, a mechanical support system consisting of an extracorporeal blood pump coupled with a gas exchange unit was attached in a right atrium-to-left atrium configuration for up to 7 days. Circuit performance, hematologic parameters, and animal hemodynamics were analyzed.

RESULTS: Six subjects underwent the chronic disease model for 56 to 71 days. Three of the subjects survived to the 7-day end-point for circulatory support. The circuit provided 2.8 (0.5) liter/min of flow compared to the native pulmonary blood flow of 3.5 (1.1) liter/min. The animals maintained physiologically balanced blood gas profile with a sweep flow of 1.2 (1.0) liter/min. Two animals freely ambulated while wearing the circuit.

CONCLUSIONS: Our novel mechanical support system provided physiologic support for a large animal model of pulmonary hypertension with right heart failure. The small footprint of the circuit and the low sweep requirement demonstrate the feasibility of this technology to enable mobile ambulatory applications.

Предметные рубрики MeSH