DOI 10.1080/00365513.2025.2479037

Медицинская библиотека · Scandinavian journal of clinical and laboratory investigation · 2025

A deep dive into four thyroglobulin immunoassays from analytical perspective.

Deveci Bulut Tuba Saadet, Serdar Muhittin Abdulkadir, Yalçın Mehmet Muhittin, Coşkun Meriç, Poyraz Aylar, Gökgöz Acar Seda, Yaz Aydın Gizem, Gülbahar Özlem

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUNDS: Serum thyroglobulin immunometric assays (sTg) are crucial for monitoring differentiated thyroid cancer (DTC) treatment. However, challenges such as anti-thyroglobulin autoantibodies (TgAb) and assay variability hinder evaluations. This study assessed four sTg methods-three second-generation (Architect, Access, Elecsys) and one first-generation (Immulite)-following Clinical and Laboratory Standards Institute (CLSI) and American Thyroid Association (ATA) guidelines.

METHODS

The study compared sTg(Architect), sTg(Access), sTg(Elecsys), and sTg(Immulite). Precision was evaluated per CLSI EP05-A3, while the lower limits of detection (LLD) were assessed using EP17-A2. Passing-Bablok and Bland-Altman analyses were conducted as per EP09c, and semi-quantitative comparisons used Kappa statistics.

RESULTS

The second-generation sTgs (Architect, Access, Elecsys) exhibited satisfactory precision (<7% coefficient of variation, CV%), unlike sTg(Immulite), which showed significant deviations and inadequate sensitivity for DTC recurrence (Limit of quantitation, LoQ = 4.59 μg/L). Second-generation sTgs had strong correlations (r > 0.884) across all concentration ranges (≤1, 1-10, >10 μg/L), with biases (slope: 1.131-2.027). sTg(Immulite) correlated well with second-generation methods for concentrations >10 μg/L (r > 0.945) but less so for <10 μg/L (r < 0.642). TgAb significantly impacted sTg(Immulite). Kappa statistics revealed strong agreement among second-generation methods (κ > 0.800) but lower concordance with sTg(Immulite), especially in TgAb(+) samples (κ: 0.562-0.653). Agreement ratios were high for second-generation methods (0.667-1.000) but variable for sTg(Immulite), particularly at lower concentrations and in TgAb(+) cases (0.097-0.727).

CONCLUSIONS

sTg(Immulite) did not meet LLD and precision criteria for DTC monitoring, facing issues with TgAb interference. Second-generation sTgs demonstrated consistent performance across all concentrations.

Для цитирования: Deveci Bulut Tuba Saadet, Serdar Muhittin Abdulkadir, Yalçın Mehmet Muhittin, Coşkun Meriç, Poyraz Aylar, Gökgöz Acar Seda, Yaz Aydın Gizem, Gülbahar Özlem. A deep dive into four thyroglobulin immunoassays from analytical perspective.. Scandinavian journal of clinical and laboratory investigation. 2025. DOI: 10.1080/00365513.2025.2479037.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.