Proof of Concept for a New Raman-Based Prototype for Noninvasive Glucose Monitoring.

Pleus Stefan, Schauer Sebastian, Jendrike Nina, Zschornack Eva, Link Manuela, Hepp Karl Dietrich, Haug Cornelia, Freckmann Guido

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Noninvasive glucose monitoring (NIGM) in diabetes is a long-sought-for technology. Among the many attempts Raman spectroscopy was considered as the most promising because of its glucose specificity. In this study, a recently developed prototype (GlucoBeam, RSP Systems A/S, Denmark) was tested in patients with type 1 diabetes to establish calibration models and to demonstrate proof of concept for this device in real use.

METHODS

The NIGM table-top prototype was used by 15 adult subjects with type 1 diabetes for up to 25 days at home and in an in-clinic setting. On each day, the subjects performed at least six measurement units throughout the day. Each measurement unit comprised two capillary blood glucose measurements, two scans with an intermittent scanning continuous glucose monitoring (CGM) system, and two NIGM measurements using the thenar of the subject's right hand.

RESULTS

Calibration models were established using data from 19 to 24 days. The remaining 3-8 days were used for independent validation. The mean absolute relative difference of the NIGM prototype was 23.6% ± 13.1% for the outpatient days, 28.2% ± 9.9% for the in-clinic day, and 26.3% ± 10.8% for the complete study. Consensus error grid analysis of the NIGM prototype for the complete study showed 93.6% of values in clinically acceptable zones A and B.

CONCLUSIONS

This proof of concept study demonstrated a practical realization of a Raman-based NIGM device, with performance on par with early-generation CGM systems. The findings will assist in further performance improvements of the device.

Для цитирования: Pleus Stefan, Schauer Sebastian, Jendrike Nina, Zschornack Eva, Link Manuela, Hepp Karl Dietrich, Haug Cornelia, Freckmann Guido. Proof of Concept for a New Raman-Based Prototype for Noninvasive Glucose Monitoring.. Journal of diabetes science and technology. 2020. DOI: 10.1177/1932296820947112.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.