DOI 10.1007/s12928-014-0263-7

Медицинская библиотека · Cardiovascular intervention and therapeutics · 2014

Stent intussusception after thromboaspiration through a platinum chrome stent: a particular case of longitudinal stent deformation.

Mila Rafael, Vignolo Gustavo, Trujillo Pedro

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

The need to improve stent deliverability has led to the development of thinner and more flexible stents. However, there is concern about decreased longitudinal strength. The number of longitudinal stent deformation reports has dramatically increased. We report a case of stent longitudinal deformation after thromboaspiration through a new generation platinum chrome bare metal stent. Images show an "intussusception effect," an extreme form of the previously described "concertina deformation," as the mechanism of shortening. Since stent technology is constantly evolving, newer devices will probably be designed to have less susceptibility to longitudinal stent deformation.

Для цитирования: Mila Rafael, Vignolo Gustavo, Trujillo Pedro. Stent intussusception after thromboaspiration through a platinum chrome stent: a particular case of longitudinal stent deformation.. Cardiovascular intervention and therapeutics. 2014. DOI: 10.1007/s12928-014-0263-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.