A radiotherapy community data-driven approach to determine which complexity metrics best predict the impact of atypical TPS beam modeling on clinical dose calculation accuracy.

Brooks Fre'Etta Mae Dayo, Glenn Mallory Carson, Hernandez Victor, Saez Jordi, Mehrens Hunter, Pollard-Larkin Julianne Marie, Howell Rebecca Maureen, Peterson Christine Burns, Nelson Christopher Lee, Clark Catharine Helen, Kry Stephen Frasier

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To quantify the impact of treatment planning system beam model parameters, based on the actual spread in radiotherapy community data, on clinical treatment plans and determine which complexity metrics best describe the impact beam modeling errors have on dose accuracy.

METHODS

Ten beam modeling parameters for a Varian accelerator were modified in RayStation to match radiotherapy community data at the 2.5, 25, 50, 75, and 97.5 percentile levels. These modifications were evaluated on 25 patient cases, including prostate, non-small cell lung, H&N, brain, and mesothelioma, generating 1,000 plan perturbations. Differences in the mean planned dose to clinical target volumes (CTV) and organs at risk (OAR) were evaluated with respect to the planned dose using the reference (50th-percentile) parameter values. Correlation between CTV dose differences, and 18 different complexity metrics were evaluated using linear regression; R-squared values were used to determine the best metric.

RESULTS

Perturbations to MLC offset and transmission parameters demonstrated the greatest changes in dose: up to 5.7% in CTVs and 16.7% for OARs. More complex clinical plans showed greater dose perturbation with atypical beam model parameters. The mean MLC Gap and Tongue & Groove index (TGi) complexity metrics best described the impact of TPS beam modeling variations on clinical dose delivery across all anatomical sites; similar, though not identical, trends between complexity and dose perturbation were observed among all sites.

CONCLUSION

Extreme values for MLC offset and MLC transmission beam modeling parameters were found to most substantially impact the dose distribution of clinical plans and careful attention should be given to these beam modeling parameters. The mean MLC Gap and TGi complexity metrics were best suited to identifying clinical plans most sensitive to beam modeling errors; this could help provide focus for clinical QA in identifying unacceptable plans.

Для цитирования: Brooks Fre'Etta Mae Dayo, Glenn Mallory Carson, Hernandez Victor, Saez Jordi, Mehrens Hunter, Pollard-Larkin Julianne Marie, Howell Rebecca Maureen, Peterson Christine Burns, Nelson Christopher Lee, Clark Catharine Helen, Kry Stephen Frasier. A radiotherapy community data-driven approach to determine which complexity metrics best predict the impact of atypical TPS beam modeling on clinical dose calculation accuracy.. Journal of applied clinical medical physics. 2024. DOI: 10.1002/acm2.14318.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.