MedElevator Войти в исследования
← Поиск публикаций

DOI 10.4244/eij-d-16-00609

Медицинская библиотека · EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology · 2017

Preclinical in vivo long-term evaluation of the novel Mitra-Spacer technology: experimental validation in the ovine model.

Peppas Athanasios, Furer Ariel, Wilson Jon, Yi GengHua, Cheng Yanping, Van Wygerden Karl, Seguin Christopher, Shibuya Masahiko, Kaluza Grzegorz L, Granada Juan F

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

AIMS: The Mitra-Spacer (Cardiosolutions, Bridgewater, MA, USA) is designed to treat mitral regurgitation by introducing a dynamic spacer that constantly adapts to the changing haemodynamic conditions during the cardiac cycle. We aimed to evaluate the performance and safety of this device in the chronic ovine model.

METHODS AND

RESULTS

Eight sheep were enrolled in this study. Through a left thoracotomy, the Mitra-Spacer was inserted via the transapical approach and advanced into the left atrium (LA) under imaging guidance. Device performance and safety were evaluated up to 90 days using fluoroscopy, echocardiography and histopathology. The volume within the balloon spacer shifted during the cardiac cycle in all cases. Seven animals survived up to 90 days for terminal imaging and tissue harvest. Echocardiography showed no change in left ventricle (LV) ejection fraction from baseline to 90 days. There were no observations of changes in LV diastolic function, pulmonary vein inflow, or tricuspid valve function. Histological analysis demonstrated no significant injury to the mitral apparatus.

CONCLUSIONS

In the healthy ovine model, Mitra-Spacer implantation was feasible and safe. At 90 days, no evidence of structural damage to the mitral apparatus or deterioration of cardiac performance was demonstrated.

Для цитирования: Peppas Athanasios, Furer Ariel, Wilson Jon, Yi GengHua, Cheng Yanping, Van Wygerden Karl, Seguin Christopher, Shibuya Masahiko, Kaluza Grzegorz L, Granada Juan F. Preclinical in vivo long-term evaluation of the novel Mitra-Spacer technology: experimental validation in the ovine model.. EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology. 2017. DOI: 10.4244/eij-d-16-00609.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.