Increasing Ablation Area Using the Arfa Ablation System: An Experimental Study in the Bovine Liver.

Ishikawa Toru, Hirosawa Hiroshi, Tanabe Yuzo, Hasegawa Iori, Sakai Nobuyuki, Honma Terasu

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND/AIM: Expanding the ablation area remains a critical challenge in the treatment of hepatocellular carcinoma. Herein, we investigated the utility of various radiofrequency ablation (RFA) protocols for increasing ablation areas, using the bovine liver as a model.

MATERIALS AND

METHODS

A bovine liver (7.2 kg) placed in a stainless-steel tray was subjected to RFA using an arfa 17-G electrode. The starting and increment wattages were varied in the linear mode. RFA was stopped after one break. The size of the discolored area, indicating thermally coagulated tissue, was measured along both the vertical and horizontal axes. The ablated area and total heat generated were calculated.

RESULTS

RFA conducted using a 30-mm needle yielded sufficient ablation areas along both the transverse and longitudinal diameters when using a low-power start at 20 W with a 5 W/min linear increase protocol. Similarly, using a 20-mm needle, a low-power start at 20 W with a 5 W/min linear increase protocol also yielded sufficient ablation areas. Although the ablation time was prolonged in both cases, the output wattage at the breakpoint and average wattage were low.

CONCLUSION

Ablation using the arfa linear mode method (i.e., starting at a low power and gradually increasing the output) can achieve a larger ablation area than standard RFA. However, this method presents the challenge of prolonged ablation time. Regardless, this approach has the potential to be a more effective procedure for yielding extensive ablation areas in clinical practice. Given the low final output levels, this finding indicates its potential utility as a minimally invasive treatment. Future clinical practices should consider strategies to address prolonged ablation times.

Для цитирования: Ishikawa Toru, Hirosawa Hiroshi, Tanabe Yuzo, Hasegawa Iori, Sakai Nobuyuki, Honma Terasu. Increasing Ablation Area Using the Arfa Ablation System: An Experimental Study in the Bovine Liver.. In vivo (Athens, Greece). 2026. DOI: 10.21873/invivo.14351.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.