MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1109/tnsre.2025.3579417

Медицинская библиотека · IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2025

Combining Multi-Canal Vestibular Implant Stimulation to Restore the Vestibulo-Ocular Reflex in 3D.

van Boxel Stan C J, Vermorken Bernd L, Volpe Benjamin, Guinand Nils, Perez-Fornos Angelica, Devocht Elke M J, van de Berg Raymond

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

INTRODUCTION

Vestibular implants targeting the semi-circular canals might offer a potential treatment approach for vestibulopathy patients. Electrical stimulation is provided to all three canals to restore gaze stabilization in all three planes of space. This study evaluated the effect of electrically stimulating multiple canals at the same time. It was hypothesized that eye responses of single-canal stimulation could predict eye responses of multi-canal stimulation, using vector summation. This might provide valuable information for future vestibular implant fitting strategies.

METHODS

Nine subjects with an investigational vestibulo-cochlear implant participated in an experiment evaluating the vestibulo-ocular reflex in response to single and multi-canal electrical stimulations. Single-canal responses were combined by vector summation to predict the multi-canal response. The prediction was compared with measured multi-canal responses. The responses were evaluated based on peak eye velocity and alignment.

RESULTS

Eye responses to multi-canal vestibular stimulation did not consistently correspond to the sum of single-canal responses. Subsets of predictions appeared within the expected range of variation, while other subsets demonstrated over- or underestimation.

CONCLUSION

Eye responses to combined multi-canal stimulation cannot be accurately predicted by summation of single-canal responses. These findings suggest that vestibular implant fitting should not rely exclusively on single-canal measurements. Instead, advanced fitting strategies that account for multi-canal interactions are likely necessary to optimize outcomes.

Для цитирования: van Boxel Stan C J, Vermorken Bernd L, Volpe Benjamin, Guinand Nils, Perez-Fornos Angelica, Devocht Elke M J, van de Berg Raymond. Combining Multi-Canal Vestibular Implant Stimulation to Restore the Vestibulo-Ocular Reflex in 3D.. IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society. 2025. DOI: 10.1109/tnsre.2025.3579417.