MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jaccas.2026.110616

Медицинская библиотека · JACC. Case reports · 2026

Algorithm-Facilitated Pacemaker-Mediated Tachycardia Mimicking Refractory Septic Shock.

Abrol Aradhya, Okpebholo Japheth, Worsnick Sarah A, Djekidel Karim, Vijayaraman Pugazhendhi

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Algorithms that minimize right ventricular pacing, such as atrioventricular (AV) hysteresis, prolong the AV delay to promote intrinsic conduction. With intact ventriculoatrial (VA) conduction, this can facilitate pacemaker-mediated tachycardia (PMT).

CASE SUMMARY: A 62-year-old woman with a dual-chamber pacemaker presented with respiratory failure and hypotension at a paced rate of 110 beats/min, treated for presumed sepsis with escalating vasopressors. Interrogation showed an upper tracking rate of 130 beats/min, AV hysteresis ON, and postventricular atrial refractory period (PVARP) (225 milliseconds) shorter than VA conduction (360 milliseconds), producing a 550-millisecond endless-loop PMT (∼109 beats/min) that escaped device detection. Disabling hysteresis and extending PVARP terminated it, allowing vasopressor weaning.

DISCUSSION: When PVARP is shorter than the VA interval, a long AV delay converts retrograde conduction into a slow re-entrant circuit that escapes device diagnostics and mimics sinus tachycardia.

TAKE-HOME MESSAGES: Suspect slow PMT in paced patients with fixed tachycardia near 100-110 beats/min when AV hysteresis is active. Disabling hysteresis and extending PVARP beyond VA conduction is diagnostic and lifesaving.

Для цитирования: Abrol Aradhya, Okpebholo Japheth, Worsnick Sarah A, Djekidel Karim, Vijayaraman Pugazhendhi. Algorithm-Facilitated Pacemaker-Mediated Tachycardia Mimicking Refractory Septic Shock.. JACC. Case reports. 2026. DOI: 10.1016/j.jaccas.2026.110616.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.