Mitroflow aortic pericardial bioprosthesis--clinical performance.

Jamieson William Robert Eric, Koerfer Reiner, Yankah Charles A, Zittermann Armin, Hayden Robert I, Ling Hilton, Hetzer Roland, Dolman William B

Оценочное исследованиеСтатьяМногоцентровое исследованиеОбзорПеревод пока не подготовлен

Аннотация

OBJECTIVE

Advancing life expectancy with the increased prevalence of aortic valve degenerative disease brings the need for an aortic bioprosthesis with excellent haemodynamic performance and comparable durability. The Mitroflow bioprosthesis has been on the worldwide market, except in the United States, since 1982, while the current model (1991) has only recently gained regulatory approval in the latter country. This study was primarily performed to determine the durability of the current Mitroflow bioprosthesis.

METHODS

The contemporary Mitroflow bioprosthesis was implanted in 381 patients in three centres. The mean age was 76.4 years (range 53-91 years) and the mean follow-up period was 5.4+/-3.4 years, a total of 2048.7 years of evaluation. Prosthesis-patient mismatch (PPM) was classified by reference effective orifice area index categories: normal > or = 0.85 cm(2) m(-2) (53.9%), mild 0.84-0.76 cm(2) m(-2) (33.9%), moderate < or = 0.75-0.66 cm(2) m(-2) (11.7%) and severe < or = 0.65 cm(2) m(-2) (0.5%).

RESULTS

The survival, at 10 years, was 39.9+/-7.9% for 50-69 years, 27.0+/-3.7% for 70-79 years and 16.6+/-4.4% for > or = 80 years (p=0.011). There was a trend (p=0.063) influencing survival for moderate-to-severe PPM. Of the independent predictors influencing survival--moderate-to-severe projected effective orifice area index (pEOAI) (Hazard Ratio (HR) 1.6, p=0.0142) and left ventricular dysfunction (ejection fraction < 35%) (HR 1.9, p=0.0193) were included. The 10-year freedom from structural valve deterioration (SVD) at explant assessing the same age groups as survival was not different (p=0.081). The 10-year actual/actuarial freedom from SVD, at explant was for > or = 60 years--94.4+/-1.4% (85.2+/-3.9%), for > or = 65 years--94.2+/-1.4% (85.0+/-4.0%), for 61-70 years--97.4+/-2.6% (95.7+/-4.3%) and for > 70 years--94.0+/-1.5% (83.2+/-4.6%).

CONCLUSIONS

The Mitroflow external mounted, pericardial aortic bioprosthesis with documented excellent haemodynamics (especially for the small aortic root), demonstrates that prosthesis-patient mismatch in moderate and severe categories can essentially be eliminated, with durability performance comparable to other heterograft (porcine and pericardial) bioprostheses.

Для цитирования: Jamieson William Robert Eric, Koerfer Reiner, Yankah Charles A, Zittermann Armin, Hayden Robert I, Ling Hilton, Hetzer Roland, Dolman William B. Mitroflow aortic pericardial bioprosthesis--clinical performance.. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2009. DOI: 10.1016/j.ejcts.2009.05.020.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.