DOI 10.1016/j.athoracsur.2006.02.035

Медицинская библиотека · The Annals of thoracic surgery · 2006

Transapical aortic valve implantation: an animal feasibility study.

Dewey Todd M, Walther Thomas, Doss Mirko, Brown David, Ryan William H, Svensson Lars, Mihaljevic Tomislav, Hambrecht Rainer, Schuler Gerhard, Wimmer-Greinecker Gerhard, Mohr Friedrich W, Mack Michael J

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Percutaneous aortic valve implantation has recently been performed in nonsurgical patients with severe aortic stenosis. Retrograde valve delivery has been problematic because of the size of the delivery system and concomitant peripheral vascular disease. We investigated a minimally invasive approach through the left ventricular apex for antegrade placement of a device-deliverable valve.

METHODS

Transapical aortic valve implantation was performed using a 23-mm equine valve mounted on a stainless steel stent in 24 swine (weight range, 35 to 45 kg). A limited or full sternotomy approach was used to access the apex of the heart. The crimped valve was introduced through a sheath in the left ventricular apex. Fluoroscopy and echocardiography were used for guidance. Deployments were performed on the beating heart either with ventricular unloading using femoral extracorporeal circulation or rapid ventricular pacing.

RESULTS

All valves were successfully delivered at the selected target site with acceptable visualization of the noncalcified aortic annulus. Valve migration occurred during eight deployments (two distal and six retrograde) secondary to persistent cardiac output, unfavorable annular anatomy, and dislodgement by the delivery catheter. Exact positioning of the nonmigrated valves at the aortic annulus was examined by necropsy of all animals at the end of the procedures. Paravalvular leak was noted in 14 of 18 (77.8%) valves remaining in situ.

CONCLUSIONS

The transapical approach was used for the successful antegrade placement of a stented valve, obviating the technical problems associated with a large delivery system transiting the peripheral vascular system. Stent design contributing to paravalvular leak remains problematic.

Для цитирования: Dewey Todd M, Walther Thomas, Doss Mirko, Brown David, Ryan William H, Svensson Lars, Mihaljevic Tomislav, Hambrecht Rainer, Schuler Gerhard, Wimmer-Greinecker Gerhard, Mohr Friedrich W, Mack Michael J. Transapical aortic valve implantation: an animal feasibility study.. The Annals of thoracic surgery. 2006. DOI: 10.1016/j.athoracsur.2006.02.035.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.